Overview

Gas mixing equipment is used for high-precision control of gas mixtures in calibration programs and for preparing gas mixtures for industrial or laboratory use. The precise dilution of various gases enables users to obtain accurate mixed gases suitable for use in various occasions. Users only need to set the target output concentration of the desired gas. The actual concentration is based on the real display during the mixing process detected by flow rate.

Product Details


principle

  • By combining various gas sensors with high-precision mass flow controllers and precise software, gas mixtures can be reduced from 100% to 1ppm.ECHO气体混合设备


purpose

  • Calibration of sensors using gas mixtures;

  • Calibration of personal gas monitoring devices;

  • Calibrate emission monitoring instruments;

  • Industrial and laboratory mixed gases;

  • Used for biotechnology, pharmacy, chemistry, and biological experiments.


advantage

  • Mixing non corrosive and corrosive gases, such as:

  • SO2,NO,NO2,CL2,H2S 等;

  • 1-4 channels;

  • High precision and repeatability;

  • Available as desktop or portable;

  • A mixture ranging from 100% to ppm;

  • Certification: The gas flow measurement laboratory is accredited by ISO/IEC 17025.


Technical SpecificationsECHO气体混合设备

  • Accuracy:+/-1% of full scale, including linearity at 15 to 25 ° C and 0.7 to 4 bar; +/-2% of full scale, including linearity from 0 to 50 ° C and 0.3 to 10 bar; Special calibration can provide+/-1% full-scale accuracy at specific temperatures and pressures;

  • Reproducibility: ± 0.25% f.s (as required ± 0.15% f.s);

  • Response time: 300 milliseconds;

  • Flow range: 0 to 10 sccm to 0 to 50 slpm; The specified flow range is the equivalent nitrogen flow rate at 760 mm Hg and 21 ° C.

  • Average reaction time: 2 seconds

  • Air pressure: (jia) 2 bar, large 34 bar;

    ECHO气体混合设备ECHO气体混合设备ECHO气体混合设备