E800 evaporative light scattering detector Sainaer

NegotiableUpdate on 07/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The E800 evaporative light scattering detector has broad compatibility with any brand of HPLC system. It has been optimized in the design of the optical path system and adopts an innovative design that is simple in structure and maintenance free, ensuring its reliable performance in terms of ease of operation, durability, and sensitivity, providing users with an efficient operating experience.

Product Details

1、 Product Features

1) Multi signal output - analog, digital signals, trigger signals, to meet different detection scenarios and requirements

2) The interior of the detection room is treated with a special coating to effectively extend the service life of the detector

3) Visual design of diversion room

4) The overall structure is compact (with a width of only 256mm), saving laboratory space

5) Adopting a fully digital mass flow controller, its performance is superior to conventional flow meters, ensuring the reliability and stability of the entire atomization, drift, and upwelling process

6) Use coaxial laser to fix the structure and avoid laser position deviation caused by vibration

7) The combination of high-sensitivity photomultiplier tube and laser ensures its high detection sensitivity

2、 Specifications

1) Light source: 650nm semiconductor laser, maximum output 30mW (light source life warranty of 35000 hours)

2) Acquisition frequency: 10Hz, 20Hz, 50Hz, 100Hz adjustable

3) Gas flow rate setting range: 0.00-5.00L/min adjustable, minimum step 0.01

4) Gas flow precision: ± 0.02L/min

5) Drift tube temperature control range: room temperature+3 ℃ to 130 ℃

6) Baseline noise: ≤ 0.035mV

7) Baseline drift: ≤ 0.35 mV/h

8) Minimum detection concentration: ≤ 0.000005g/mL (cholesterol/methanol solution, with column detection)

9) Quantitative repeatability: RSD ≤ 2.0%