- Phone
-
Address
906, Building 1, Yuzhou City Plaza, Lane 49, Chuanshan Street, Pudong New Area, Shanghai
Shanghai Taoshi Industrial Development Co., Ltd
906, Building 1, Yuzhou City Plaza, Lane 49, Chuanshan Street, Pudong New Area, Shanghai
E+H Proline Promag electromagnetic flowmeter
The specific parameters of the E+H Proline Promag H 200 electromagnetic flowmeter are as follows:
• Nominal caliber range: DN2... 25 (1/12... 1 ')
• Material of liquid receiving components: PFA lining, electrode material 1.4435 (316L), Alloy C22, 2.4602 (UNS N06022) alloy, tantalum, platinum
• Measurement variables: volumetric flow rate, mass flow rate
Maximum measurement error: Volume flow rate is ± 0.5% O.R. ± 2mm/s (± 0.08in/s)
• Measurement range: 0.06dm ³... 300m ³/h (0.015... 80gal/min)
Maximum process pressure: PN40, Cl.150, 20K
• Medium temperature range: -20...+150 ° C (-4...+302 ° F)
• Environmental temperature range: -40...+60 ° C (-40...+140 ° F)
Sensor housing material: 1.4301 (304), corrosion-resistant
• Material of transmitter housing: with aluminum alloy AlSi10Mg coating
• Protection level: IP66/67, Type 4X
E+H Proline Promag electromagnetic flowmeter
• Display/Operation: Four line backlit display, touch key operation (external operation); Set up through on-site display screens and debugging software; Provide separate display units
• Output: 4... 20mA HART (passive), pulse/frequency/switch output (passive)
• Input: None
• Digital communication: HART, PROFIBUS PA, FOUNDATION Fieldbus
• Power supply: 18... 35VDC (4... 20mA HART, with or without pulse/frequency/switch output)
Explosion proof certification: ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC, TIIS
Product safety: CE certification, C-Tick certification
• Functional safety: Functional safety follows the IEC 61508 standard and is allowed to be used in safety applications specified in the IEC 61511 standard
• Metrological certification and certification: Perform calibration on the certification calibration device (compliant with ISO/IEC 17025 standard); Heartbeat Technology meets the traceability verification requirements of ISO 9001:2008 standard section 7.6a (T Ü V certification)
• Pressure certification and certification: CRN
• Material certificate: 3.1 Material certificate