- Phone
-
Address
B6-101, Zhongnan High tech Intelligent Manufacturing Industrial Park, No. 10 Shuangma Street, Yuetang District, Xiangtan City, Hunan Province
Hunan Zhenhua Analytical Instrument Co., Ltd
B6-101, Zhongnan High tech Intelligent Manufacturing Industrial Park, No. 10 Shuangma Street, Yuetang District, Xiangtan City, Hunan Province
STD-BDielectric constant and dielectric loss testerIt consists of a high-frequency Q-meter, an S96 testing device (fixture), data acquisition and tan δ automatic measurement controls, an LKI-1 inductor group, and a liquid cup (optional for testing liquid materials). It is designed and manufactured in accordance with the national standards GB/T 1409-2006, GB/T 1693-2007, American standard ASTM D150, and International Electrotechnical Commission IEC60250. The system provides an automatic measurement solution for the high-frequency dielectric loss tangent (tan δ) and dielectric constant (ε) of insulation materials.
STD-BDielectric constant and dielectric loss testerMain technical indicators
host |
STD-B high-frequencyQwatch |
test device |
S916 |
LKI-1Type inductor group(9A) |
√ |
Liquid cup |
√ |
1. STD-BHost parameters:
Function Name: |
STD-B |
Range of signal sourceDDSDigital synthesis signal |
10KHZ-110MHz |
Signal source frequency coverage ratio |
11000:1 |
Signal source frequency accuracy 6Significant number of digits |
3×10-5 ±1a character |
Qmeasurement range |
1-1000automatic/manual range |
Qresolution |
resolution0.1,4Significant number of digits |
QMeasurement work error |
<5% |
sampling precision |
11BIT |
Inductance measurement 4Significant number of digits,resolution0.1nH |
1nH-8.4H ,resolution0.1nH |
Inductance measurement error |
<5% |
Tuning capacitor |
Main capacitor30-540pF |
capacitanceDirect measurement range |
1pF~2.5uF |
Tuning capacitance error,resolution |
±1pFor<1% ,0.1pF |
Resonance point search |
auto-scan |
QQualified preset range |
5-1000Sound and light prompts |
QRange switching |
automatic/manual |
LCDDisplay parameters |
F,L,C,Q,Lt,Ctwait |
Residual inductance and test lead inductance |
have |
Direct measurement and display function for large capacitance values |
The measured value can reach2.5uF |
Dielectric loss coefficient |
precision 3/10000 |
Minimum dielectric loss |
1/10000 |
dielectric constant |
precision one thousandth |
Material testing thickness |
0.1mm-10mm |
2. Dielectric constant and dielectric loss testing fixtureS916:
a)4Digital explicit adjustment micro rod; Fixture loss tangent value ≤4×10-4 (1MHz).
b)Polar size≧38mmAdjustable range of pole spacing: ≥15mm;Fixture plug spacing:38mm±0.01mm;Tangent value of fixture loss: ≤4×10-4 (1MHz); Micrometer resolution: ≤0.001mm.
c)Liquid cup:diameter48mmdepth7mm,can be paired withS916Swap the lower pole pieces for use.
3. Inductance groupLKI-1:5mH,2.5mH,1mH, 250MH,100MH,25MH, 10MH,2.5MH,0.5MH.