Technical Parameter
|
·Measurement range
|
Minimum 0~350m, maximum 0~2000m
|
|
|
·Measurement accuracy
|
(Comprehensive errors include linearity, hysteresis, and repeatability)
|
|
|
|
Simulated 0.2% FS; Intelligent 0.2% FS, 0.1% FS, 0.5% FS
|
|
|
·Long term stability
|
Better than 0.03% per year
|
|
|
·Allowable medium temperature
|
-20℃~90℃
|
|
|
·Allow ambient temperature
|
-40℃~85℃
|
|
|
·Allow storage temperature
|
-40℃~85℃
|
|
|
·Temperature influence
|
0.2%, 0.1%, 0.075% accuracy level -20 ℃~80 ℃, every 10K+0.1%, other temperature ranges every 10K ± 0.2%
|
|
|
|
0.05% accuracy level, every 10K ± 0.02%
|
|
|
·Material of the part in contact with the medium
|
|
|
·silicon piezoresistive sensor
|
Corrugated film 316L (special requirements agreed upon)
|
|
|
·Measuring sensor housing
|
Stainless steel 316L
|
|
|
·Invest in extension cables
|
Plastic PE
|
|
|
·Sensor counterweight
|
Stainless steel 304 (special requirements agreed upon)
|
|
|
·Material of non-contact parts with the medium
|
|
|
·Junction box
|
ABS plastic IP66/IP67
|
|
|
·Process Connection Standard
|
Nickel plating on carbon steel
|
|
|
·Working voltage
|
(12.5~30V DC)
|
|
|
·Signal output
|
Analog type: 2-wire system 4-20mA
|
|
|
|
Intelligent two-wire 4-20mA with HART communication protocol
|
|
|
·Cable entry hole
|
Internal thread M20 × 1.5 (PVC junction box)
|
|
|
·Analog cable
|
Three core instrument cable intelligent cable: It is recommended to use shielded twisted pair cable for wiring
|
|
|
·Allow environmental humidity
|