Chlorine hydrogen analyzer
It can meet the challenge of detecting chlorine and hydrogen in all nitrogen and oxygen (so-called inert gases) in chlorine production plants. It uses an inert material thermal conductivity sensor (thermal conductivity meter) equipped with two detection chambers, controlled by a microprocessor for signal processing, making it a highly reliable and accurate process monitoring device.
K6500 can be used to monitor the electrolysis and liquid chlorine sections of all types of chlor alkali plants. We can provide turnkey projects for it.
Chlorine hydrogen analyzeradvantage
Sensors are non consumable parts and are not easily damaged.
The interchangeability of sensors reduces downtime to the shortest possible time
The sensor is not affected by fluctuations in sample gas and environmental pressure, and is particularly stable.
If the sample trachea is contaminated, it is particularly easy to clean
Use alternative gas for verification, with a long verification cycle.
Chlorine hydrogen analyzerTechnical specifications
range
0-100% Cl2- display up to 0.1%, resolution up to 0.04%
0-5% H2- Display to 0.01%, resolution to 0.004%
0-10% H2- Display up to 0.01%, resolution up to 0.008%
Please specify the range of hydrogen gas when placing an order
stability
Better than 1% of full scale within the entire operating temperature range
Better than 1% of full scale per month
precision
± 1% or ± 2% of full scale - dependent on verification method
reproducibility
Better than 1% of full scale
Sample gas flow rate
*Operating flow rate between 100-300 ml/min
sample gas temperature
0 ℃ -100 ℃ (non agglomerating)
sample gas pressure
The work pressure is between 0.9-1.1 bar (ambient atmospheric pressure)
Zui maximum absolute pressure 2 bar
Sample gas connection: inlet and outlet
What is the suitable outer diameter? PFA double clamp compression joint for pipes
ambient temperature
Sensor:+5 to+55 ℃
Monitor: -10 to+45 ℃
display
Instrument: 4-row LCD
Reactor controller: 4-digit LED
Reactor malfunction: Front panel LED
analog output
The isolated linear 4-20mA: 20mA point corresponding to the hydrogen range can correspond to any value between 20% and 100% of the hydrogen range.
Isolation linear 4-20mA corresponding to chlorine gas range.
Maximum load of 600 ohms
Alarm relay output
The alarm output of the four-way voltage free relay is detailed as follows. Each relay is a conversion contact, with a maximum capacity of 48V, 0.5A, A.C. or D.C
Instrument status
Reactor heater status: ± 10 ℃ of the set temperature value
Programmable Cl2 concentration alarm
Programmable H2 concentration alarm
response speed
T90=30 seconds
Power requirements
110Vac or 230Vac ± 10%, maximum power consumption of 80VA.
controller
Installation: Wall mounted casing
Dimensions: Height 231mm, Width 225mm, Depth 225mm
(Excluding cable sleeves), refer to Figure 1
Material: polycarbonate shell, with front cover and terminal cover, equipped with suitable cable sleeve, protection level IP54.
Sensor/Reaction Chamber
Installation: Wall mounted PVC panel
Dimensions: Height 90mm, Width 240mm, Depth 150mm
Excluding the heat shield, refer to Figure 2.
Sensor to controller: up to 30 meters away. As standard configuration, a 2-meter cable is provided for connecting to the sensor. Each sensor provided has its own electrical identification module, which can be used in conjunction with all types of controllers.