Chlorine Oxygen Analysis System

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Online chlorine oxygen analyzer, using imported oxygen analyzer, with special corrosion-resistant gas chamber and pipeline, to meet all oxygen analysis applications in acidic gas environments. Suitable for oxygen content analysis in high concentration acidic environments such as chlorine gas, hydrogen chloride, hydrogen sulfide, and sulfur dioxide. Application scenarios: chemical plants, petrochemical plants, chlor alkali plants, fluorine chemical industry, sponge titanium process, chlorination process for titanium dioxide, methane chloride process, etc.

Product Details

gas analyzerUE610 - O2It uses the principle of paramagnetic cross pressure to measure the oxygen content in corrosive gases, and can be used in general safety areas and special explosion-proof areas.


Technical specifications


Measuring range:

Detection range

Measure small values

Measure large values

Measurement range

(including zero correction)

characteristic

Control Panel


The measurement results show that


Electromagnetic Compatibility


installation orientation

size

weight



Measurement response:

Noise (electronic time constant)1s, Range2

A small amount of process0.5%(The small quantity is indicated on the nameplate)

zero drift<Small range on the label0,5% / 3moon

Measurement drift<Each range's0,5% / 3moon

reproducibility<Each range's1%

Influencing variables:

ambient temperature < 0.5% / 10 K

sample gas pressure

No pressure compensation: range based< 2%1%Changes in pressure

Pressure compensation: range based< 0.2%1%Changes in pressure

Residual gas Zero drift corresponding to magnetism or diamagnetic drift of residual gas

(See operating manual)AD 55 0012

Sample gas flow rate<The small possible range on the nameplate1%Within the allowed flow range

Within the scope, there are0,1 l/minChanges in traffic volume


Electronic input and output:

analog output 0 / 2 / 4~20mAFloating load750

relay output 6Different contacts can be freely set, such as range recognition: AC/DC 24 V / 1Afloat

Analog input 2Group, used for0/2/4~ 20 mACorrection of external pressure sensors and residual gas effects (correction of cross effects)

Binary input 6, used for24VFloating, freely adjustable, such as range selection

serial interface RS 485

Optional parts Additional electronic accessories:8A binary input and8Binary output, such as automatic calibration trigger

Climate conditions:

Allowable ambient temperature-30~ +70 °CStorage and transportation environment

+5 to +45 °CWorking environment

Allow humidity Annual average< 90% RHDuring storage and transportation

protection levelIP 20 (EN 60529)

Multi range automatic switching

0~ 2% v/vor5% v/v O2

100% v/v O2

When the standard gas is passed, in0~100%v/vof

Any zero point between them is possible (see reference gas table)

linear

Liquid crystal display,LEDBacklight and contrast control, soft buttons, numeric keys, function keys

4½ -bit,Depending on the selected measurement range; Optional decimal numbers

part,

Front panel vertical

19”, 4 HU = 177 x 483 x 476 mm

about20Kg























































power supply

AC 200 ~ 240 V, 48~63 Hz(Rated range: 180 V to 264 V)

energy loss about35 VA

Inlet conditions:

sample gas pressure ±0,5~±1,5Ba (flexible hose);

±0,5~±3 Ba (pipeline)

Sample gas flow rate20~ 60 l/h (0,3 ~ 1 l/min.)

sample gas temperature0~50 °C

sample gas humidity< 90% RH

Response time:

warm-up time at room temperature< 30 min

Display delay time(T901.5~3.5sDepending on the model

Attenuation (electronic time constant)0~100s, can be set

Death time about0,5 ~2,5Depending on the model