- Phone
-
Address
No. 860 Xinyang Road, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
Shanghai Sanwei Scientific Instrument Co., Ltd
No. 860 Xinyang Road, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
Three provides for scienceIndustrial control and reactor matchingHastelloy high-pressure parallel flow pump, Hastelloy high-pressure constant flow pump, Hastelloy high-pressure plunger pumpSupports industrial networks such as Profinet, Ethercat, Ethernet/IP, Modbus/TCP, etc., making it easy to connect and communicate with various automation devices in industrial applicationsProvide integrated supporting services for chemical instruments and equipment such as reactors, microreactors, and synthesis devices.Our full range of high-pressure parallel flow pump products can be matched with microchannel reactors, high-pressure reactors, petroleum development evaluation experiments, catalytic reactions in petrochemicals, polymerization reactions, food, pharmaceuticals, liquid chromatography, separation and purification, atomic energy science, coal, metallurgy, environmental science, process equipment, and experimental equipment for accurate and trace transportation of various liquids.
The flow range of Hastelloy series products includes: 0.00110ml/min; 0.0150ml/min; 0.01100ml/min; 0.01200ml/min; 0.01300ml/min; 0.01600ml/min; 0.11000ml/min。

H1110 Hastelloy high-pressure metering pump (1000ml pump head, 10 Mpa) Product features:
Accurate delivery of hardware functions
Multi point flow calibration: achieving high accuracy and high repeatability in fluid transport across the entire range
Flow pulse suppression: Cam curve compensation and flow pulse electronic suppression effectively control pressure pulses
Pressure setting correction: protection pressure setting of the flow path, zero point value correction to eliminate environmental interference
Piston cleaning function: After cleaning the plunger, reduce the wear of the sealing ring and extend the service life of the pump
Weight reduction and weighing control: customized reactor and chemical plant inlet weighing closed-loop control system
Integration of communication function system
Multiple communication modes: one-to-one control, one to many control, many to many control
Multiple communication ports:RS232/485/422TheUSBWired/wireless network
Multiple communication protocols: customizableModbus RTU/ASCIITheProfibusDP、Profinet, etc
Support system integration: integrationSCADA/DCS/PLCControl system and configuration application
Analog digital control: analog input control, switch start stop control
Software function extension application
Transport unit switching: Switch the flow rate unit and select to transport by volume or weight
Flow rate variation control: constant, linear, gradient flow rate delivery, can be derived in any combination
Real time display of readings: real-time drawing of flow rate and pressure curves, display of program running progress bar
Data Record Export: Real time recording of flow and pressure data, exported in tabular and graphical form
Method save and call: Save and call multiple flow rate programs, control convenient switching between methods
Calibration and maintenance function: Customize flow calibration to meet specific operating conditions and restore factory parameters
Remote control function: optional remote software, remote real-time operation and monitoring of fluid equipment
Technical parameters of H1110 Hastelloy high-pressure metering pump:
serial number |
description |
indicator |
1 |
Infusion method |
Dual plunger parallel mode, floating plunger design |
2 |
flow range |
0.1-1000.0ml/min |
3 |
increment |
0.1ml/min |
4 |
Traffic accuracy |
± 0.5% |
5 |
Traffic repeatability |
≤ 0.5% |
6 |
pressure range |
≤ 10Mpa |
7 |
pressure pulsation |
≤ 0.5Mpa |
8 |
Flow path material |
Hastelloy, ruby PTFE、 ceramic |
9 |
Pipeline connection |
1/8 'Hastelloy alloy pipeline connection |
10 |
Display parameters |
256 * 64 dot LCD display, self luminous display screen |
11 |
Communication function |
RS232,USB, RS485 ; Support three company protocols and Modbus protocol |
12 |
power supply |
85 ~ 264VAC,50Hz |
13 |
size |
550×420×222 mm3 |