Applicability of Siemens gas analyzer 7MB2335-0NH06-3AA1

NegotiableUpdate on 05/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Siemens gas analyzer 7MB2335-0NH06-3AA1 is suitable for automatic device calibration using ambient air (depending on the measured component); The cycle time can be adjusted within the range of 0 (1)... 24 hours. The gas analyzer ULTRAMAT 23 is used to measure one infrared component and a 19 inch pull-out device for measuring oxygen and hydrogen sulfide. It is installed in the control cabinet to measure the gas interface 6mm tube.

Product Details

Applicability of Siemens gas analyzer 7MB2335-0NH06-3AA1Use ambient air (depending on the tested component) for automatic device calibration; The cycle time can be adjusted within the range of 0 (1)... 24 hours. The gas analyzer ULTRAMAT 23 is used to measure one infrared component and oxygen and hydrogen sulfide. The 19 inch pull-out device is installed in the control cabinet to measure the gas interface 6mm tube. There is no internal measurement gas pump to measure the SO2 concentration range. Parallel measurement of SO2 and NO2 with UV photometer. Minimum measurement sample: 0 to 500 vpm. Maximum measurement range: 0 to 2500 vpm. No electrochemical oxygen sensor. No H2S sensor. Auxiliary energy: 230VAC, 50Hz. Operating software, documentation in English.
西门子气体分析仪7MB2335-0NH06-3AA1适用性

Siemens gas analyzer 7MB2335-0NH06-3AA1Ordering Instructions

When measuring certain components, special selection rules must be followed.


Measurement of component nitrous oxide


7MB2335, 7MB2337, and 7MB2338

(Application: Silicon wafer production)


The measurement range is from 0 to 100/500 ppm (MB label "E")

Can only be used to measure nitrous oxide in ultra pure gases

7MB2337 and 7MB2338

(Application: Measurement according to Kyoto Protocol requirements)


The measurement range is from 0 to 500/5000 vpm (MB label "Y")

Simultaneously measuring carbon dioxide is required to correct for cross interference

7MB2337- * CP * 0- * SY * or

7MB2338- * DC * 0- * SY * (including nitric oxide measurement)


7MB2337 and 7MB2338

(Application with paramagnetic oxygen measurement unit and independent gas path)


7MB2337-4**80-**** - Z + C11

7MB2337-5**80-**** - Z + C11


7MB2338-4**80-**** - Z + C11

7MB2338-5**80-**** - Z + C11


Measure component SF6


7MB2335, 7MB2337, and 7MB2338

(Application: Silicon wafer production)


The measurement range is 0 to 500/2500 ppm (MB label "H")

Can only be used to measure SF6 gas in ultra pure gas

Calibration interval (MCERTs type 7MB2335, 7MB2337, 7MB2338)

ULTRAMAT 23, 19 'rack mounted and portable versions

General Information


Measure components

Maximum 4

range

2/Measurement components

display

LCD display with adjustable LED backlight contrast; Function building; 80 characters (4 lines/20 characters)

operating position

Front wall panel, vertical

compliance

CE certification (EN 61000-6-2, EN 61000-6-4)

Design, shell


weight

About 10 kg

Protection level, 19 'rack and desktop models

  • 7MB2335, 7MB2337, and 7MB2338:
    IP20, Compliant with standard EN 60529

  • 7MB2355, 7MB2357, and 7MB2358:
    IP40, Compliant with standard EN 60529

Electrical performance


EMC anti-interference (electromagnetic compatibility)
Safety Extra Low Voltage (SELV) with Safety Isolation

Meets the requirements of NAMUR NE21 or EN 61326-1 standards

auxiliary power supply

  • 100 V AC, +10%/-15%, 50 Hz

  • 120 V AC, +10%/-15%, 50 Hz

  • 200 V AC, +10%/-15%, 50 Hz

  • 230 V AC, +10%/-15%, 50 Hz

  • 100 V AC, +10%/-15%, 60 Hz

  • 120 V AC, +10%/-15%, 60 Hz

  • 230 V AC, +10%/-15%, 60 Hz

power consumption

About 60 VA

Electrical input and output


Analog output

By component, 0/2/4... 20 mA, NAMUR, floating point, maximum load 750 Ω

relay output

8. With conversion contacts that can be freely set, such as range recognition;

24 V AC/DC/1 A, Floating point, no spark

Digital input

3. Designed for 24V without voltage

  • pump

  • Automatic Calibration

  • synchronization

serial interface

RS 485