89.157.017 Oxygen electrode
CNA-89.157.017 Oxygen Electrode
87.213.050 oxygen electrode
CNA-87.213.050 Oxygen Electrode
09182=A=1000 oxygen electrode
Z09182=A=1000 oxygen electrode
PM-DL5PPB oxygen electrode

Various oxygen electrodes

Match SWAN instrument

Match HACH instrument

Match Rosemount oxygen meter

OXY500 oxygen electrode
Our company's research and developmentoxygen electrodeCan be used in conjunction with SWAN oxygen meters, HACH oxygen meters, Rosemount oxygen meters, and most oxygen meters on the market. Easy to use, can replace oxygen membrane and electrolyte or directly replace membrane head, simple operation, long quality and service life.
Installation methods: circulation type, sinking type, or oblique insertion type.
Application areas: wastewater treatment, thermal power generation, aquaculture, pharmaceuticals, high-temperature fermentation, chemical industry, papermaking, semiconductor and other industries.
Main technical references:
Measurement range: 0.00~20.00PPm (mg/L)
Response time: (T98%, 25 ℃)
Temperature range: 0-130 ℃
Ordering selection: 9500M is designed for constant measurement (ppm level);
9500P is used for trace measurement (ppb level);
9500U is used for measuring the oxygen content in high-temperature sterilization;
Filling and membrane replacement
Turn off the power.
1. Unscrew the electrode cap and pour out the remaining filling solution.
2. Take a new electrode cap and fill it with new electrolyte.
There is a groove on the electrode thread, which allows excess air and electrolyte to be discharged. Hold the electrode vertically, with the electrode probe facing downwards and the groove on the upper side.
3. Gently tighten the electrode cap to allow excess electrolyte to be discharged without generating too much pressure inside the electrode.
4. Clean the electrode and dry the sensing film with a soft tissue.
Connect the power supply. Let the electrode run in the air for at least 30 minutes, not less than 1 hour. Afterwards, perform air calibration and reinstall the electrode after completing the calibration.
oxygen electrodeMaintenance of
Sometimes it is necessary to replace the electrolyte and oxygen permeable membrane. Warning: The filling solution is alkaline and corrosive, containing less than 1% potassium hydroxide. Do not inhale, wear protective goggles and gloves when filling, and avoid contact with clothing. Once it splashes into the eyes, please rinse immediately with clean water and contact a doctor. Short term contact with the skin is harmless, but it still needs to be rinsed with plenty of water