Welcome Customer !

Membership

Help

Beijing Zhichen Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Beijing Zhichen Technology Co., Ltd

  • E-mail

    250939901@qq.com

  • Phone

    15710019877

  • Address

    No. 18 Jianshe Road, Liangxiang, Fangshan District, Beijing

Contact Now

chromatograph

NegotiableUpdate on 01/27
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The GC-902 series is a practical gas chromatograph with a single detector configuration. The instrument can be equipped with hydrogen flame ionization detector (FID), thermal conductivity cell detector (TCD), electron capture detector (ECD), etc. The instrument is equipped with an intelligent rear door, which can perform constant temperature and 9-level programmed heating operations. The instrument adopts a color touch screen operation interface, which is convenient and intuitive to operate. It has a compact structure, beautiful appearance, good reliability, and high cost-effectiveness. It can be widely used in production and research departments in industries such as petrochemicals, pesticides, environmental protection, and food inspection.
Product Details

GC-902 series gas phasechromatographPractical gas phase configuration for single detectorchromatographThe instrument can be equipped with hydrogen flame ionization detector (FID), thermal conductivity cell detector (TCD), electron capture detector (ECD), etc. The instrument is equipped with an intelligent rear door, which can perform constant temperature and 9-level programmed heating operations. The instrument adopts a color touch screen operation interface, which is convenient and intuitive to operate. It has a compact structure, beautiful appearance, good reliability, and high cost-effectiveness. It can be widely used in production and research departments in industries such as petrochemicals, pesticides, environmental protection, and food inspection.

Technical Specifications:

Column box:

Temperature control range: Room temperature+3 ℃~399 ℃

Control accuracy: ± 0.1 ℃

Temperature accuracy: ± 3%

Cooling rate: from 250 ℃ to 50 ℃ for about 8 minutes

Optional detector:

FID detector:

Detection limit: D ≤ 1x10 'g/s (n-hexadecane/isooctane)

Noise: N ≤ 2x10-14A

Drift: dr ≤ 2x10-3A/30min

Linear range: ≥ 106

TCD detector:

Sensitivity: ≥ 4000mv. ml/mg (benzene/toluene)

Noise: N ≤ 0.02mv

Drift: dr ≤ 0.2mv/30min

Linear range: ≥ 104

ECD detector

Detection limit: D ≤ 1X10-1 ° g/ml (γ -666/iso octane)

Noise: N ≤ 0.04mv

Drift: dr ≤ 0.3mv/30min

Linear range: ≥ 103