Trace sulfur analyzerPerformance characteristics:
Pick up limit (based on S in H2S): 0.01mg/M3
Undiluted measurement range: 0.01-50 (0.1-500) mgS/M3 (calculated as S in H2S).
Data repeatability: Using bottled H2S standard gas for 7 consecutive injections, the square root error of the analysis results is less than 5%.
Column temperature: adjustable from room temperature to 200 ℃± 0.2 ℃.
Permeation source temperature: adjustable from room temperature to 200 ℃± 0.1 ℃.
Detector temperature: adjustable from room temperature to 300 ℃± 1 ℃.
Converter temperature: adjustable from room temperature to 1000 ℃± 5 ℃.
High voltage power supply: adjustable from -10V to -900V.
Trace sulfur analyzerApplication scope:
A. Coke oven gas, synthetic gas, dry natural gas, biogas, sulfur recovery equipment, etc
B. Organic phosphorus residue detection in food and water quality
C. Sulfides in the ambient air (odorous gases such as wastewater discharge from aquaculture and chemical plants)
D. Combustion of exhaust emissions, treatment of automobile exhaust emissions (sulfur dioxide detection)
E. Sulfides in high-purity gases (food grade carbon dioxide)
F. Coking benzene, thiophene in petroleum benzene (0.05ppm)
G. Sulfides in automotive hydrogen fuel cells