Overview

The micro differential pressure transmitter uses differential capacitance as the detection principle to form a micro differential pressure capacitive transmitter, with an input pressure of 0-0.1-1.0KPa and an output of 4-20mA DC analog signal. The product features advanced design principles, complete variety and specifications, and easy installation and use.

Product Details

One product overview

ZY-1151DR micro differential pressure transmitter is used to measure small differential pressures such as furnace pressure, and then convert them into 4-20mA DC signal output. The intelligent micro differential pressure transmitter can also communicate with the HART controller for setting, monitoring, or forming an on-site monitoring system with the upper computer. ZY-1151DR on-site adjustable intelligent differential pressure transmitter is a new product developed by our company according to on-site requirements. It can be operated without a manual controller and can achieve on-site zero adjustment, configuration, and other operations through button pressing.

II Product performance specifications
(Zero benchmark calibration range, under reference conditions, filled with silicone oil and 316 L stainless steel isolation diaphragm.). )

1.1. Reference accuracy
1.1.1. Digital and intelligent: ± 0.2% calibration range
1.1.2. Simulation, linear: ± 0.5% calibration range
1.2. stability
1.2.1. Digital and intelligent: 6 months, ± 0.2% URL
1.2.2. Simulation, linear: 6 months, ± 0.5% URL
1.3. Environmental temperature impact
1.3.1. Numbers and Intelligence:
Zero point error: ± 0.2% URL/56 ℃
Overall error: ± (0.2% URL+0.18% calibration range)/56 ℃
1.3.2. Simulation, Linear
Zero point error: ± 0.5% URL/56 ℃
Overall error: ± (0.5% URL+0.5% calibration range)/56 ℃
1.4. Static pressure effect
The zero point error is ± 0.5% URL. It can be corrected online by resetting the zero.
1.5. vibration effect
In any axial direction, the vibration effect at 200Hz is ± 0.5% URL/g
1.6. Power supply impact: less than ± 0.005% output range/volt.
1.7. Load impact:
There is no load impact unless there is a change in the power supply voltage.
1.8. Electromagnetic interference/radio frequency interference (EMI/RFI impact)
When the field strength reaches 30V/M from 20 to 1000MHz, the output drift is less than ± 0.1% of the range
1.9. Installation location impact
The zero drift is at most ± 0.25kPa. All zero drift can be corrected; Has no impact on the measurement range.

2. Functional specifications
2.1. Measurement range: Differential pressure: 0-0.125-1.5KPa
Static pressure: 1, 4, 6.9MPa
2.2. Zero point and range
2.2.1. Digital and intelligent: can be adjusted with local range and zero buttons, or remotely adjusted with a HART controller

2.2.2. Simulation, linear: range and zero point continuously adjustable
2.3. Zero point positive and negative transfer
When zero point negative shift occurs, the lower limit of the range is greater than or equal to - URL; When the zero point is shifted forward, the upper limit of the range is less than or equal to+URL. Verify that the range is greater than or equal to the minimum range.
2.4. output
Digital and intelligent:
4~20mA DC, Users can choose linear or square root output, and the digital process variable is superimposed on a 4-20mA DC signal, which can be used by upper computers using the HART protocol.
Simulation, linear: 4-20mA DC, linear with process pressure.
2.5. Damping time constant
Digital and intelligent: The time constant is adjustable, increasing by 0.1 seconds from a minimum to 16.0 seconds
Simulation, linear: The time constant can be adjusted from a minimum of 1.67 seconds.
2.6. Upper limit of ambient temperature
Digital and intelligent: -40 to 85 ℃
Simulation, linear: -40 to 93 ℃
Equipped with LCD meter head: -30 to 60 ℃
2.7. Upper limit of process temperature
Silicone oil filling: -40 to 104 ℃
Fill with inert liquid: -18 to 71 ℃
2.8. Upper limit of storage temperature
Digital and intelligent: -51 to 85 ℃
Simulation, linear: -51 to 121 ℃
Equipped with LCD meter head: -40 to 70 ℃
2.9. ambient humidity
0-100% relative humidity
2.10. Static pressure and overpressure upper limit
Applying a pressure of 0 to 6.9 MPa on either side of the transmitter will not cause damage. Meets performance specifications when operating within a static pressure range of 3.45kPa to 6.9MPa.
2.11. Volume change amount
Less than 0.16cm3

III Product Selection

ZY-1151DR

Micro differential pressure transmitter

code

Range KPa

2

0-0.125~1.5KPa

code

output

F

4-20mA

code

Structural materials

Flanges and fittings

Exhaust/Drain Valve

Isolation Diaphragm

Fill with liquid

22

316 stainless steel

316 stainless steel

316 stainless steel

silicone oil

code

Working pressure MPa

A-

1

B-

4 (Special 6.9)

code

option

M1

0-100% linear indicator table

M2

LED display table

M3

LCD display table

B1

Pipe bending bracket

B2

Board mounted bending bracket

B3

Pipe mounted flat bracket

D1

The side relief valve is located at the upper part of the pressure chamber

D2

The side relief valve is located at the lower part of the pressure chamber

No note

1/2NPT taper pipe threaded joint

C2

T-shaped threaded joint M20 × 1.5, with a ball cone connector welded with a rear end of a Ø 14 pressure pipe

d

Explosion proof type d Ⅱ BT4

i

Intrinsic safety type ia Ⅱ CT6

s

Stainless steel three valve group