Introduction:
ZLS-2Refrigeration type circulating water meterIt is a low-temperature liquid circulation equipment that uses mechanical refrigeration. It has the function of providing low-temperature liquid and low-temperature water bath. Combined with moisture analyzer, rotary evaporator, vacuum freeze-drying oven, circulating water multi-purpose vacuum pump, magnetic stirrer and other instruments, various low-temperature chemical reaction operations and drug storage can be carried out.
The digital display of this instrument is intuitive, simple and clear, and easy to operate.
The host and cooling liquid storage tank are integrated for easy installation and use, and are not affected by noise and temperature.
There are protections for high and low pressure, temperature, overcurrent, overvoltage, etc. The equipment will work normally in a safe environment when operated strictly according to requirements.
This device is particularly suitable for chemical, biological, and physical experiments that require maintaining low temperatures and working at room temperature. It is an essential equipment for laboratories in the fields of medicine and health, chemical industry, food industry, metallurgical industry, colleges and universities, research institutions, genetic engineering, polymer engineering, etc.
Installation and Attention:
The distance between the host and its surroundings must not be less than0.5Mi!
Placed on a solid foundation to prevent intense vibrations and shaking.
Cannot be used in corrosive environments.
Cannot be used in environments with flammable or explosive gases.
Do not work in a dusty environment. Please regularly clean the dust on the condenser to ensure that the refrigeration system is working properly. Environmental temperature<30℃。
The host casing must be connected to the ground wire, and the ground wire conductor must not be less than2.5mm2.
Cannot overload operation.
Frequent starting of this machine will affect the lifespan of the motor, and the number of starts per hour should be less than6Next time.
When not in use for a long time, the water pump should be powered on before starting up to test whether it is running, whether the instrument display is normal, whether the compressor can work normally after a three minute delay, and then the cooling system should be started up to make the system work normally.
Setting method:
SM-101Instructions for setting up microcomputer controller:
1Main functions
(1)Temperature display
(2)Single cooling
(3)Temperature upper and lower limit control
(4)Automatic reset and storage function
(5)Compressor startup delay protection for three minutes
(6)The controller automatically ensures that the upper limit is greater than the lower limit
(7)Temperature correction is used to correct temperature measurement errors.
2.technical performance
(1)Temperature display range+48℃~-48℃
(2)Temperature setting range ±48Set arbitrarily within ℃
(3)Temperature measurement accuracy ±1℃
(4)Resolution1℃
(5)Temperature correction ±5℃ can be set arbitrarily.
3.Set upper and lower limits (there are two ways to operate the instrument, depending on which one you are using, please follow the instructions in the table)
(1)Pressing the "Select" button will display the upper or lower temperature limit on the digital display, and then pressing the "Add" or "Subtract" button will change the set value. After setting the lower and upper temperature limits, you must press the "Select" button to return to the displayed temperature measurement state (automatic reset)12The set value can only be saved in seconds.
(2)Press the selection button, when the temperature lower limit indicator light flashes, press the plus or minus button to set the temperature lower limit (compressor stop), then press the selection button again, the temperature return difference (<16)When the indicator light is on, press the plus key to set the upper temperature limit (compressor on), for example: lower temperature limit-15℃+Temperature difference5℃=-10℃(-15℃Shutdown, -10℃Power on)
Press the selection button again, and when the display screen flashes, press the plus or minus button to set the temperature correction.
4Compressor startup delay protection
There is a "compressor shutdown timer" inside the controller, which starts counting when the compressor stops. The next time the compressor is started, check the timer first. If it has been three minutes, start the compressor immediately. If it is less than three minutes, wait until it has been three minutes before starting the compressor. In addition, the controller will not start the compressor within three minutes of being powered on, to prevent damage to the compressor due to frequent starts.
Suspend delay: When the power is first turned on, hold down the "add" button and do not release it5Just a second.
5Fault display
(1)The temperature probe is faulty, open or disconnected, and the digital display will show“E1”.
(2)Temperature probe malfunction or short circuit in the wiring, temperature above90℃, temperature below-43At ℃, the digital display shows“E2”(Appears)“E1”Or“E2”Line breaks or sensors).
Operation Instructions:
1.Add non freezing liquid into the storage tank (in-50Non freezing liquid at ℃, approximately from the edge of the tank100mm.
2.Press the power button on the device, and the display panel will show the cooling outlet temperature.
3.Set the desired temperature according to the setting method.
4.Press the cooling switch, and after a three minute delay, the compressor starts working.
5.Open each valve, press the switch of the circulation pump, and the circulation pump will start working to make the temperature in the storage tank uniform.
6.After a period of time, when the displayed temperature reaches the desired temperature, other work can begin. The device can automatically control temperature.
7.After the work is completed, turn off the cooling switch and stop2~3After minutes, turn off the circulation pump switch, and finally, pull down the main power switch.
8.ambient temperature<5℃When shutting down for a long time, it is necessary to drain the non freezing liquid (especially water with high components) of the circulating pump and the layout heat exchanger to prevent the circulating liquid from freezing and damaging the equipment.
Attention: If the non freezing liquid used contains water absorbing components such as ethylene glycol, please add the non freezing liquid in a timely manner. Otherwise, the heat exchanger may freeze due to the increase in freezing point of the non freezing liquid, blocking the system's expansion and cracking equipment, affecting the cooling effect of the whole machine, and in severe cases, the equipment may be damaged.
Main technical parameters:
indicator parameter
project
power supply 220V 50HZ
Compressor power, power supply 1horse 2.8A
refrigeration capacity 700W~2KW
Circulating pump performance Power:100Welectric current:0.5A
Traffic:80L/minLift:19rice
Scope of application -30degree~+30℃
Intelligent digital temperature control temperature range: ±50℃ accuracy: ±2℃
Storage tank capacity 13L
ambient temperature <<30℃
Dimensions of the sink 300×300(mm)
host size 575×440×795(mm)
High pressure pumps can be equipped according to user requirements.
After-sales service
Users can use it under normal conditions from the date of shipment12If the product malfunctions due to manufacturing quality reasons within a month, our factory is responsible for free repair or replacement. Our factory will charge repair costs for damages caused by improper use.
Telephone:0530-2481019
Fax0530-2481210
Address: Linbu Industrial Zone, Juancheng, Shandong8number
URL:http://www.hualw.com