E-mail
huanuodianzi@126.com
-
Phone
13839220101
-
Address
Chengbei Industrial Zone, Hebi City, Henan Province
Huanuo Electronic Technology Co., Ltd
huanuodianzi@126.com
13839220101
Chengbei Industrial Zone, Hebi City, Henan Province
1Oxygen bomb automatic lifting, automatic inner bucket constant water volume, automatic control of inner bucket water temperature regulation, automatic completion of experiments.
2Advanced compressor refrigeration technology, completely unaffected by environmental temperature changes, ensures that the temperature difference between the inner and outer cylinders of the instrument meets national standards. Can work continuously for a long time.
3The calorimeter supports two ignition methods: ignition wire and cotton thread
4Configure compressor constant temperature water tank,Observe whether the inner cylinder is short of water
5Stainless steel vacuum inner cylinder. The mixing system adopts German imported motors.
6Fast testing speed and testing cycle≤8min(Quick method)≤15min(national standardGB/T213-2003).
7Thermal capacity stability<0.2%precision<0.1%temperature resolution0.0001K
8This product has excellent performance and reliability even when operating in harsh environments.
9Compact structure, beautiful appearance, easy installation and maintenance, low failure rate.
10Compact in size, extensively manufactured using molds, with high precision.
11The repeatability and reproducibility of the heat generation test are superior to the national standardGB/T212-2003The requirements.
12High degree of automation, automatic utilization of the water volume in the built-in fixed container, automatic control of the temperature difference between the inner and outer buckets of the instrument, and automatic completion of the entire experimental process. Can be connected to an electronic balance.
13useWindowsXPThe operating system enables multiple controls on one machine, with no impact on each other's testing and high software stability.
14The data processing function is rich, and users can easily query historical experimental data, current day data, parallel sample data, etc.
15. Can be connected to an electronic balance to automatically input the weight of the sample
Temperature measurement range:0-40℃
refined secret Degree:≤0.1%
divide distinguish Rate:0.0001K
Power supply voltage:AC220±10% 50Hz
Accuracy: superior to national standards
Thermal capacity stability: Changes in thermal capacity within three months ≤0.10%; Change in heat capacity within one year ≤0.15%GB/T483-2007General Regulations for Coal Analysis Experimental Methods
GB/T 213-2008Method for determining the calorific value of coal
GB/T 384-1981Method for determination of calorific value of petroleum products
GB/T 30727-2014Method for determination of calorific value of solid biomass fuels
GB/ T14402—2007Combustion performance of building materials and products - Determination of calorific value of combustion
GB/T 30991-2014General Technical Conditions for Intelligent Oxygen Bomb Calorimeter
JC/T 1005-2006Method for determining the calorific value of cement black raw materials
JJG 672-2001Verification Regulations for Oxygen Bomb Calorimeter
ASTM D5865-2013Method for determination of calorific value of coal and coke
ASTM D5865-2007Standard Test Method for Total Calorific Value of Coal and Coke
ISO 1928-2009Solid Mineral Fuel-Measurement of Total Value and Calculation of Net Heat Value Using a Bomb Calorimeter
CEN/TS 14918Method for determination of calorific value of solid biofuels
BS EN 15400-2011Solid Recycling Fuel-Heat generation test
IS: 1350-1970Methods for determination of coal and coke
GJB5891.29-2006Test Methods for Explosive Ordnance Agents29Partial "Constant temperature method for combustion and explosion heat determination"