I. Overview
XOUV-III type ultraviolet radiation analyzerIt is a versatile 254nm ultraviolet radiation system primarily used for crosslinking nucleic acids onto membranes. It can also be used for DNA cutting in agar gel, RecA mutation screening, partial restriction endonuclease digestion produced by thymic dimer, and UV sterilization to eliminate PCR pollution. It also has applications in UV sterilization, polymer UV treatment, and other fields.
2、 Instrument features:
To fix the nucleotide on the membrane, the traditional method is to place the membrane in a vacuum oven at 80 ° C for 2 hours, and then irradiate it with 254nm ultraviolet light for a few seconds on a spectrophotometer.
UV irradiation can increase the hybridization signal by 5-10 times compared to traditional baking methods. Due to the configuration of a UV energy programming system (Joules/cm2) in the UV diplomatic instrument, the time integrator continuously detects the irradiation of ultraviolet light.
When the energy absorption value reaches the set value, the irradiation will automatically stop. So regardless of the strength of the ultraviolet light source or the time difference, the irradiation cycle has good repeatability.
Touch screen display and operation
◎ UV occlusion window
◎ Large stainless steel UV exposure room
◎ Power outage information storage
IIIXOUV-III type ultraviolet radiation analyzerTechnical Indicators:
◎ UV wavelength: 254nm (additional 312nm and 365nm lamps are provided according to user requirements)
◎ UV light source: 5 10W light tubes
9 exposure energy settings and save
Set and save 9 exposure times
◎ Manual setting of UV exposure energy
◎ Manual setting of UV exposure time
◎ Exposure energy measurement range: 0-99.99J (joules)
◎ Exposure time measurement range: 0-99.9 minutes
4、 Structural features:
◎ Push pull suction door, well sealed
◎ External dimensions Length: 45cm Width: 36cm Height: 32cm
Exposure Room Size: Length: 34cm, Width: 26cm, Height: 15cm
◎ Weight 10Kg
5、 Usage conditions:
◎ Environmental temperature: 5-35 ℃
◎ Relative humidity: ≤ 80%
◎ Power supply voltage: single-phase AC 220V fluctuation range ≤ 20%
◎ Power frequency: 50Hz
6、 Software system usage instructions:
6.1 Main Screen
Ø Joule: Click on "Joule" to enter the Joule settings interface
Ø Timer: Click on "Timer" to enter the time setting interface
Ø Settings: Click on "Settings" to enter the wavelength selection interface
Main screen
6.2 Joule setting interface
⑥ |
⑤ |
④ |
③ |
② |
① |
Joule setting interface
① PV: Actual Joule Value Display
② SV: Joule setting value, click in to set the Joule value
③ Home: Button to return to the main screen
④ Stop: Stop Joule Run Key
⑤ Start: Start Joule Run Key
⑥ 0-8 can set 9 exposure energies and save them
6.3 Time Setting Interface
⑥ |
⑤ |
④ |
③ |
② |
① |
Time setting interface
① PV: Actual running time display
② SV: Run time setting value, click in to set the time
③ Home: Button to return to the main screen
④ Stop: Stop time running key
⑤ Start: Start time running key
⑥ 0-8 can set 9 exposure times and save them
6.4 Wavelength Selection
Select the corresponding wavelength based on the actual wavelength of the lamp tube on the UV Vis analyzer, and click the confirm button

Wavelength selection interface (standard UV lamp UV254nm: 5 pieces)