XOUV-III type ultraviolet radiation analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The XOUV-III ultraviolet radiation analyzer is a versatile 254nm ultraviolet radiation system primarily used for crosslinking nucleic acids onto membranes. It can also be used for DNA cutting in agar gel, RecA mutation screening, partial restriction endonuclease digestion produced by thymic dimer, and UV sterilization to eliminate PCR pollution. It also has applications in UV sterilization, polymer UV treatment, and other fields.

Product Details

f9b271ee576501b1fe604263b171941

I. Overview

XOUV-III type ultraviolet radiation analyzerIt is a versatile 254nm ultraviolet radiation system primarily used for crosslinking nucleic acids onto membranes. It can also be used for DNA cutting in agar gel, RecA mutation screening, partial restriction endonuclease digestion produced by thymic dimer, and UV sterilization to eliminate PCR pollution. It also has applications in UV sterilization, polymer UV treatment, and other fields.

2、 Instrument features:

To fix the nucleotide on the membrane, the traditional method is to place the membrane in a vacuum oven at 80 ° C for 2 hours, and then irradiate it with 254nm ultraviolet light for a few seconds on a spectrophotometer.

UV irradiation can increase the hybridization signal by 5-10 times compared to traditional baking methods. Due to the configuration of a UV energy programming system (Joules/cm2) in the UV diplomatic instrument, the time integrator continuously detects the irradiation of ultraviolet light.

When the energy absorption value reaches the set value, the irradiation will automatically stop. So regardless of the strength of the ultraviolet light source or the time difference, the irradiation cycle has good repeatability.

Touch screen display and operation

◎ UV occlusion window

◎ Large stainless steel UV exposure room

◎ Power outage information storage

IIIXOUV-III type ultraviolet radiation analyzerTechnical Indicators:

◎ UV wavelength: 254nm (additional 312nm and 365nm lamps are provided according to user requirements)

◎ UV light source: 5 10W light tubes

9 exposure energy settings and save

Set and save 9 exposure times

◎ Manual setting of UV exposure energy

◎ Manual setting of UV exposure time

◎ Exposure energy measurement range: 0-99.99J (joules)

◎ Exposure time measurement range: 0-99.9 minutes

4、 Structural features:

◎ Push pull suction door, well sealed

◎ External dimensions Length: 45cm Width: 36cm Height: 32cm

Exposure Room Size: Length: 34cm, Width: 26cm, Height: 15cm

◎ Weight 10Kg

5、 Usage conditions:

◎ Environmental temperature: 5-35 ℃

◎ Relative humidity: ≤ 80%

◎ Power supply voltage: single-phase AC 220V fluctuation range ≤ 20%

◎ Power frequency: 50Hz

6、 Software system usage instructions:

6.1 Main Screen

Ø Joule: Click on "Joule" to enter the Joule settings interface

Ø Timer: Click on "Timer" to enter the time setting interface

Ø Settings: Click on "Settings" to enter the wavelength selection interface

e88e1d5e3098698df3daa5200e30a2f

Main screen

6.2 Joule setting interface

⑥

⑤

④

③

②

①

191e9e8b91052a410cf8e7d3fe1c62c

Joule setting interface

① PV: Actual Joule Value Display

② SV: Joule setting value, click in to set the Joule value

③ Home: Button to return to the main screen

④ Stop: Stop Joule Run Key

⑤ Start: Start Joule Run Key

⑥ 0-8 can set 9 exposure energies and save them

6.3 Time Setting Interface

⑥

⑤

④

③

②

①

e4864b357dba317d3806aa7d2327f93

Time setting interface

① PV: Actual running time display

② SV: Run time setting value, click in to set the time

③ Home: Button to return to the main screen

④ Stop: Stop time running key

⑤ Start: Start time running key

⑥ 0-8 can set 9 exposure times and save them

6.4 Wavelength Selection

Select the corresponding wavelength based on the actual wavelength of the lamp tube on the UV Vis analyzer, and click the confirm button

f1ce6486f7fdd140d87a273053e5d7c

Wavelength selection interface (standard UV lamp UV254nm: 5 pieces)