Suitable for industries such as electricity, coal, metallurgy, petrochemicals, environmental protection, cement, papermaking, geological exploration, and research institutions to determine the moisture, ash content, and volatile matter of coal, coke, or other organic matter, and calculate their fixed carbon and calorific value. It can also analyze fly ash, ash residue, coke, and carbon content. The automatic industrial analyzer adopts the standard GB/T212-2008 "Methods for Industrial Analysis of Coal".
The main features of XDGY-3000 automatic industrial analyzer are:
1Standardized process and accurate results: The testing process, conditions, and results for moisture, ash content, and volatile matter all meet the national standard requirements and can meet the testing requirements for all types of coal without calibration, and can be used for arbitration analysis.
2Fast testing speed: Using a single imported balance, a multi furnace structure, and a constant temperature drying device, the three indicators of moisture, ash content, and volatile matter can be combined or measured separately. At the same time, testing the three indicators of 19 samples can be completed within 150 minutes. Especially suitable for customers who are not familiar with the sample. The fewer samples, the shorter the time.
3High degree of automation: RS485 on-site industrial control bus is used for communication, high-precision imported electronic balance is integrated into the instrument, combined with a set of pneumatic controlled automatic weighing mechanism, using thermogravimetric analysis method to automatically weigh, send, process data, calculate results, print and store reports, etc
4Easy to operate and safe: There is no need to remove or place crucible covers, send samples, or take samples during the testing process, avoiding the risk of high temperature radiation and burns.
5High cost-effectiveness: Using a single testing host to achieve precise measurement of three industrial analysis indicators in accordance with national standards, with high cost-effectiveness and complete functionality.
6Dynamic addition of samples: During the testing process, samples can be temporarily added to adapt to special situations where the user's sample is uncertain.
Technical parameters of automatic industrial analyzer:
1Sample weight: 0.5-1.5g
2Maximum working temperature: 1000 ℃
3Temperature control accuracy: ± 1 ℃
4Sample quantity: 1-19 samples of water, ash, and volatile are made simultaneously
5Test temperature: 105 ℃ (moisture content)
815℃ (ash content)
900℃ (volatile matter)
6Testing time: 19 samples, three indicators testing time ≤ 150 minutes
7Precision: GB/T212-2008 "Industrial Analysis Methods for Coal".
8Accuracy: within the allowable deviation range of the standard sample
9Gas requirements:
Oxygen: purity 99.5%, pressure reduced to 0.1Mpa
Nitrogen: purity 99.5%, pressure reduced to 0.1Mpa
Pressure reducer: High end 0-25Mpa, Low end 0-0.4Mpa
10Power: ≤ 4kW
11Power supply: (220 ± 22) VAC, (50 ± 1) Hz
12Dimensions: 620X530X760mm
12Weight: 65Kg