The rapid automatic hydrogen analyzer is mainly used to determine the content of hydrogen element in coal and other solid materials. The entire testing process is controlled by a microcontroller, which corrects and processes the test data. The digital display shows the milligrams or percentages of hydrogen in the measured substance, and the final result is output by a printer. The measurement time for each sample is about 10-15 minutes, with fast speed, accurate results, simple operation, and high degree of automation. It is currently the most ideal hydrogen measuring instrument in the laboratory. This instrument can also measure carbon through the weighing absorption method, and its accuracy and precision meet the requirements of the national standard GB/T476-2001.
Technical specifications and parameters of XDCQ-5 rapid automatic hydrogen analyzer
Measurement range: Hydrogen 0-20%, Carbon 1-100%
Heating rate: 25 ℃/min, 30 minutes or rising to 800 ℃
Measurement time: approximately 10-15 minutes
Heating control: 25K/minute, rise to 600 ℃ in 30 minutes; Linearity:+ 0.1%
Allowable error: Hydrogen: ≤ 0.15% (same laboratory)
Carbon: ≤ 0.50%
Power supply: AC220V ± 10%, 50Hz
Temperature control: 800 ℃± 10 ℃, 300 ℃± 10 ℃
Linearity: ± 0.1%
Sample: Particle size<0.2mm
Weight: Approximately 65mg
Accuracy and precision: comply with the requirements of the national standard GB/T476-2001.