Mainly used to determine the content of carbon and hydrogen elements in coal and other solid materials. The testing process is controlled by an integrated microcontroller, which corrects and processes the test data, displays the milligrams or percentage of hydrogen content in the measured substance digitally, and outputs the results through a printer. This instrument measures quickly, provides accurate results, is easy to operate, and has a high degree of automation. The number displays the milligrams of hydrogen in the measured substance and automatically prints the dry basis content of hydrogen. The appearance design is novel, easy to operate, reduces the usage space, and is more convenient for maintenance. This instrument can also measure the carbon content in coal and other solid materials. Measurement accuracy meets the requirements of GB/T476-2001
Technical specifications and parameters of an integrated rapid automatic hydrogen analyzer
Measurement range: Hydrogen 0-20%, Carbon 1-100%
Heating rate: It can rise to 850 ℃ in 50 minutes
Measurement time: approximately 10-15 minutes
Measurement accuracy: meets the requirements of GB/T476-2001
Instrument stability linearity: ± 0.1%
Heating control: 25K/minute, rise to 600 ℃ in 30 minutes; Linearity:+0.1%
Allowable error: Hydrogen: ≤ 0.15% (same laboratory) Carbon: ≤ 0.50%
Power supply: AC220V ± 10%, 50Hz
Temperature control: 800 ℃± 10 ℃, 300 ℃± 10 ℃
Linearity: ± 0.1%
Sample: Particle size<0.2mm, approximately 65mg