Welcome Customer !

Membership

Help

Chuzhou Patex Instrument Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Chuzhou Patex Instrument Technology Co., Ltd

  • E-mail

  • Phone

  • Address

    Economic Development Zone No. 6

Contact Now

X-ray fluorescence sulfur analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The X-ray fluorescence sulfur analyzer is developed and manufactured to meet the needs of sulfur content detection in oil products. It adopts the principle of energy dispersion to measure the total sulfur mass percentage content of crude oil, petroleum, heavy oil, diesel, kerosene, gasoline, and naphtha; Measure the total sulfur content in coal chemical products, such as primary benzene; Measurement of total sulfur or sulfide content in other liquid or solid fine powder samples. It complies with the relevant requirements of national standards GB/T17040 and GB11140, as well as the stricter requirements of the American national standard ASTMD4294-02.

Product Details

X-ray fluorescence sulfur analyzerIt is a specialized analyzer designed based on energy dispersive X-ray fluorescence spectroscopy. The principle, design structure, instrument components, analytical method operation process, error and reproducibility of the analyzer all comply with GB/T 17040-2019 Determination of sulfur content in oil and oil products.

X-ray fluorescence sulfur analyzerMain purpose:

Measure the sulfur content of crude oil, crude oil, heavy oil, diesel, kerosene, and naphtha;

Measure the total sulfur content in coal chemical products, such as primary benzene;

Measurement of sulfur content in other liquid or solid fine powder samples such as oil coke.

X-ray fluorescence sulfur analyzerInstrument features:

Intelligent electromechanical integrated microcomputer design, 8-inch wide video and capacitive touch screen (1024 * 768) display

◆ Wide variety of detection, wide detection range, fast analysis speed, and low consumption of standard samples

The instrument has automatic peak stabilization function and automatic diagnostic function, monitoring the working status and electrical parameters of the instrument

◆ One time Mylar film sample cup to avoid cross contamination; Sample cup making is convenient and fast

The sample stage is accurately positioned, making it easy to place samples and replace leak proof oil films, thus avoiding the possibility of contamination of the detection system

The instrument has a large data storage capacity, with 4096 content analysis results, 8192 counting measurement data, and 16 instrument calibration result data

Thermal printer automatic printing detection results

The instrument automatically selects the working curve without the need for user pre selection

Users can choose the measurement result unit, ppm or (m/m)%

◆ Safe radiation protection measures, with no ionizing radiation harm to individuals or groups

Microcoulometric sulfur analyzerTechnical Indicators:

1. Sulfur measurement range: 0.0017% to 5% to 99% optional;

2. Measurement time: 60-180 seconds;

3. Single sample automatic measurement, measurement repetition times: 1, 2, 3, 5, 10 can be set arbitrarily, and the average value and standard deviation will be given at the end of the measurement;

4. Repeatability (r):<0.4347 X0.6446 (X=average of two repeated test results);

5. Reproducibility (R):<1.9182 X0.6446 (X=average of two single and independent test results);

6. Sample size: 5ml to 6ml;

7. The instrument can store 10 calibration curves;

8. Working conditions: Temperature: 10-30 ℃; Relative humidity: ≤ 85% (30 ℃); Power supply: AC100-220V/50Hz; Rated power: 50W;

9. Dimensions and weight: 460mm * 260mm * 260mm; 11kg。