Microcoulometric sulfur chlorine analyzerOverview:
The KVD-600 microcoulometric titrator is a new generation instrument developed based on the KVD-200 microcoulometric titrator, which adopts the principle of dynamic microcoulometric titration. The sample to be tested undergoes oxidation-reduction reaction through combustion in the cracking tube, and is carried into the titration cell by the carrier gas to convert into titratable ions, which react with the electrically generated I3- or Ag+. By measuring the amount of I3- or Ag+consumed during the electrolytic titration process, the total sulfur or total chlorine content of the sample can be obtained according to Faraday's law.
Scope of application: Used for the analysis of total sulfur or total chlorine content in samples in the production, scientific research, and monitoring fields of petroleum, petrochemicals, pharmaceuticals, health, environmental protection, coal, geology, metallurgy, commercial inspection, quality inspection, and schools.
Compliant with standards:
SH/T0253 Determination of sulfur content in light petroleum products (electric method)
SH/T0222 Determination of Total Sulfur Content in Liquefied Petroleum Gas (Electricity Method)
GB/T11061.4 Determination of total sulfur content in natural gas (oxidation microcoulometric method)
GB/T3208 Determination of sulfur content in benzene products (electric method)
SH/T1147 Determination of Sulfur Content in Industrial Aromatic Hydrocarbons (Microcoulometric Method)
GB/T18612-2011 Determination of Organic Chlorine Content in Crude Oil (Microcoulometric Method)
SH/T1757 Determination of Chlorine Content in Industrial Aromatic Hydrocarbons (Microcoulometric Method)
Microcoulometric sulfur chlorine analyzerTechnical parameters:
Measurement range: Sulfur: 0.2ppm~10000ppm~percentage content Chlorine: 0.3ppm~10000ppm~percentage content
Measurable sample status: solid, liquid, gas (with corresponding injector optional, standard for liquid injection)
Recommended injection volume: Solid:>=5mg Liquid: 1 to 100ul Gas: 2 to 5ml
Temperature control range and accuracy: Room temperature to 1050 ℃, ± 1 ℃
Repetitive error: X < 1.0ppmCv ≤± 0.2ppm
1.0ppm≤X≤10ppmCv≤±10%
X>10ppmCv≤±5%
Working power supply: AC220V, 3KW
Gain adjustment: automatic adjustment
Sampling resistor: automatic adjustment
Bias voltage: continuously adjustable from 0 to 500mV, real-time monitoring
Instrument completeness: host (amplifier), temperature control (cracking) system, stirrer, sampler (liquid), computer (optional), operating software, electrolytic cell, quartz tube, etc
Features:
*The instrument adopts USB communication technology, which makes the connection control more reliable and has a lower failure rate!
*The instrument separates sulfur and chlorine according to their high and low content, and changes the gain and sampling resistance from manual adjustment to automatic adjustment!
*The instrument has achieved full-time monitoring of bias voltage. Easy for users to operate!
*The instrument adopts air cooling for heat dissipation, the fan automatically switches on and off, and no one needs to be on duty to cool down when turned off!
*The instrument adopts an independent temperature control system. Adopting Yudian's fully automatic temperature controller, the failure rate is close to zero!
*The electrolytic cell adopts 304 stainless steel electrode caps, which can be used for a long time and avoid corrosion by the electrolyte!
*Windows operating platform, human-machine direct dialogue, implementing foolproof operation!
*Computer controls the entire process of analysis, data processing, etc., displaying the working status and peak shape of the entire process. Parameters, data, and peak shapes can be saved or printed as needed!
*The instrument adopts internationally popular circuits and imported components, reducing instrument noise and improving detection sensitivity!
*The instrument measurement results are accurate and stable!
User supporting facilities:
Workbench: 1500mmX600mm Power supply: 220V AC not less than 4KW
Gas source: ordinary oxygen>99.9% ordinary nitrogen>99.9%
Chemical reagent: (KI) analytical grade 100g
(NaN3) analytical purity 100g (only used for analyzing sulfur content)
Ice acetic acid (HAC) high-grade pure 500Ml
5L of purified water