E-mail
LabStone@vip.163.com
- Phone
-
Address
6th Floor, Building A1, No. 2-6 Jinbao Road, Baiyun Lake Street, Baiyun District, Guangzhou City
Guangzhou Xitang Electromechanical Technology Co., Ltd
LabStone@vip.163.com
6th Floor, Building A1, No. 2-6 Jinbao Road, Baiyun Lake Street, Baiyun District, Guangzhou City
STW-8-080-F
Function
Water vapor transmission performance test - water vapor transmission capacity, transmission rate, and transmission coefficient.
Films, sheets, boards, containers, and building materials made of plastic, fabric, leather, metal, and other materials.

Working principle
Balance weighing: weight reduction method, can be placed in both upright and inverted cups.
Place pure water in the sample cup; Seal the sample cup with the sample and place it inside the testing box; Control the temperature, humidity, and wind speed of the test box; The humidity difference inside and outside the sample cup drives water vapor to penetrate the sample, resulting in a decrease in the mass of the sample cup; Regularly weigh the sample cup and calculate performance parameters such as the water vapor transmission rate of the sample.
Compliant with standards
Chinese Pharmacopoeia (Part IV) - Determination of Water Vapor Transmittance GB/T 1037、YBB 00092003 ASTM E96、ISO 2528、GB/T 17146 GB/T 12704 GB 19082 GB/T 21332、QB/T 2414、TPPI T464、ASTM D1653、YY0471.2、DIN 53122。
FEATURES
TECHNICAL INDEX
|
Name |
parameter |
name |
parameter |
|
measuring range |
Diaphragm: 0.05-10000 g/m2•24h; Container: 0.0002-30 g/pkg • 24h |
test error |
Diaphragm: 0.05 g/m2•24h; Container: 0.0002 g/pkg • 24h |
|
Scale range |
210 g |
Scale division of the balance |
0. 1 mg |
|
temperature range |
10-50 ℃ |
Temperature control error |
±0.1 ℃ |
|
Humidity range |
5% - 50% RH |
Moisture control error |
±1% RH |
|
Wind speed |
0.5-2.5 m/s (optional 0.03-0.5 m/s) |
Mpa |
>0.3 MPa (self provided) |
|
test area |
33 cm2 |
specimen thickness |
<3 mm |
|
Container quality |
<180 g |
Container height |
<150 mm |
|
Power supply |
AC220V±10% 50Hz |
Power |
<2 kW |
|
Model differentiation |
Model number |
||
|
STW-8/080-F |
|||
|
Number of samples |
8 |
||
|
Host size |
873(L)x700(B)x688(H) |
||
|
Net weight of the host |
126 kg |
||
Standard configurationTest host, computer, air dryer, professional testing software, breathable cup (number of workstations), calibration weight, sampler.
Optional configurationSoftware module of GMP computerized system, wax sealed sample cup.
Bring your ownCompressed air.