Water vapor transmission performance test - water vapor transmission capacity, transmission rate, transmission coefficient, wet flow density, and moisture permeability.
Suitable for testing the water vapor transmission performance of building materials, thin sheets, and other samples.

Working principle
Balance weighing: integrating weight loss and weight gain methods.
Drying agent in the sample cup (weight gain method) or placing water (weight loss method); Seal the sample cup with the sample and place it inside the testing box; Control the temperature, humidity, and wind speed of the test box; The humidity difference inside and outside the sample cup drives water vapor to penetrate the sample, resulting in an increase or decrease in the mass of the sample cup; Regularly weigh the sample cup and calculate performance parameters such as the water vapor transmission rate of the sample.
Execution standards
ASTM E96, GB/T 17146, GB/T 21332, QB/T 2411, etc.
FEATURES
TECHNICAL INDEX
|
Name |
Parameters |
Name |
Parameters |
||
|
measuring range |
permeance |
0.05~10,000g/m2•24h |
test error |
0.05 g/ m2•24h |
|
|
|
Wet flow density |
5.8*10-10~1.2*10-4Kg/s. m2 |
|
5.8*10-10 Kg/s. m2 |
|
|
Scale range |
620 g |
Scale division of the balance |
1 mg |
||
|
temperature range |
10-50 ℃ |
Temperature control error |
±0.1 ℃ |
||
|
Humidity range |
5% - 95% |
Moisture control error |
±1% RH |
||
|
Wind speed |
0.02~0.5m/s (optional 0.02~0.3m/s) |
Mpa |
>0.3 MPa (self provided) |
||
|
test area |
50cm2 (customized for other areas) |
sample thickness |
≤150 mm |
||
|
Power supply |
AC220V±10% 50Hz |
Power |
<2000 W |
||
|
Model differentiation |
Model number |
||||
|
|
W-B-31-F-1 |
W-B-61-F-1 |
|||
|
Number of samples |
3 |
6 |
|||
|
Host size |
873(L)× 700(B)× 688(H) |
||||
|
Net weight of the host |
133kg |
135 kg |
|||
Standard configuration:Test host, computer, air dryer, professional testing software, wax sealed breathable cup (number of workstations), calibration weight, sampler.
Optional configuration:Software module of GMP computerized system.
Self provided:Compressed air, distilled water, desiccant.