- Phone
-
Address
No. 85 Xigang North Road, Haigang District, Qinhuangdao City
Qinhuangdao Xunling Technology Development Co., Ltd
No. 85 Xigang North Road, Haigang District, Qinhuangdao City
Water hardness online analyzer economical DUROMAT IQ imported
Standard measurement range:
Total hardness: 0.22-534 mg/L CaCO ∝; 0.002-5.3mmol/L
Total alkalinity: 5.34-401.8 mg/L; 0.11-8.04mmol/L
Measurement principle:
Online monitoring of water hardness/alkalinity through automatic titration colorimetric method.
Typical applications:
Online monitoring of hardness/alkalinity of water quality such as softened water, boiler water, pure water, drinking water, and circulating water.
Application industry:
Industries such as water plants, pharmaceuticals, steel, swimming pools, beer and beverage, semiconductors, solar energy, electricity, heat, and chemicals.
Instrument features:
• Automatic online analysis, simple operation menu
• Monitoring limits and programmable output control
The same analyzer can measure total hardness or alkalinity
• Built in self calibration and self diagnostic functions
Low maintenance and low reagent consumption
• Compact design
LCD graphic display
Relay output limits and fault error messages
Analog signal output 0/4-20mA: measurement value or status information
• Can be continuously measured or measured with time intervals (5-360min) or initiated by external switch signals
Built in SD card, measurement results, device fault information, and device firmware information are all recorded on the SD card
Optical components without condensation
• Software updates via SD card
technical indicators
Measurement parameters: total hardness, total alkalinity
Measurement method: Titration colorimetric method
Standard measurement range: Total hardness: 0.22-534 mg/L CaCO ∝; 0.002-5.3mmol/L
Total alkalinity: 5.34-401.8 mg/L; 0.11-8.04 mmol/L
Measurement accuracy: ± 5% of the upper limit value of the selected reagent
Repeatability: ± 2.5% of the upper limit of the selected reagent
Reagent consumption: approximately 0.05-0.5 ml/analysis, depending on the measured water hardness/alkalinity
Measurement time: Approximately 3 minutes, depending on the hardness/alkalinity of the water and the set flushing time
Analysis frequency: MAX 10000 analysis/500ml low water hardness reagent.
Reagent expiration date: 2 years after leaving the factory, (<25 ° C, away from light)
Input/Output: 2 sets of programmable relays output maximum. 250 VAC 4A
• Limit value alarm
• Error messages
1 set of 0/4-20 mA signal output with a maximum of 750 Ω
1 set of signal input
External control analysis begins
• Flow control
Water sample consumption: about 1L/time (water pressure 2Bar). The specific water sample consumption varies depending on the inlet pressure and internal flushing time
Power supply voltage: 85-264 VAC (50-60 Hz) or 24V AC/DC
Power consumption: 25VA (during operation) 3.5VA (during standby)
Protection level: IP43 without casing or IP56 standard casing, optional IP65
Storage temperature: 0 ° C-45 ° C
Environmental temperature: 10 ° C-45 ° C
Water temperature: 5 ° C-40 ° C
Air humidity: 20-90% RH (without ice or condensation)
Inlet pressure: 0.5 bar -5 bar (recommended 1-2 bar)
Water sample requirements: clear, colorless, no suspended solids, no bubbles, pH:4-10, Iron:<3ppm, copper:<0.2 ppm, aluminum:<0.1 ppm; Manganese:<0.2 ppm, acid capacity: KS 4.3<5 mmol/l
Display units: ° f, ° dH, ppmCaCO3, mmol/l, ° e
Installation method: Wall mounted
Equipment size: Without protective box 280 x 360 x 113 mm (W x H x D)
300 x 390 x 130 mm (W x H x D) with protective box
Equipment weight: Approximately 1.9KG without protective box, approximately 3.8KG with protective box
Water hardness online analyzer economical DUROMAT IQ imported