Walk in high and low temperature environment laboratoryA high-precision constant temperature and humidity laboratory refers to a laboratory with a relatively dense distribution of temperature and humidity, similar to a regular constant temperature and humidity chamber5~10Degree error, andLiuqin TechnologyThe error distribution of the assembled constant temperature and humidity laboratory is positive and negative2Within the range of degrees, the fluctuation is positive or negative0.5Within degrees,Temperature analysis accuracy:0.01℃,湿度解析精度:0.1%R.H,Temperature control accuracy :±0.2℃,Humidity control accuracy:±2%R.H.
They are suitable for testing various performance indicators of aerospace products, information electronic instruments and meters, electrical and electronic products, materials, various industrial products or electronic components in high and low temperature constant temperature or humid and hot constant humidity environments. High precision constant temperature and humidity laboratory | The control part of the high-precision constant temperature and humidity laboratory is designed on the right side of the equipment, equipped with temperature controllers, control switches, etc. It is easy to operate, simple, and easy to maintain. thislaboratorySelect high-quality materials and use current processing equipment to shape them.
Walk in high and low temperature environment laboratoryHigh precision constant temperature and humidity laboratoryProtection system:
1Overall equipment overheating;
2Fan overheating protection
3The overall equipment is out of phase/Reverse phase;
4Overload of refrigeration system;
5Refrigeration unit overpressure;
6Water shortage protection for humidification system;
7Overall equipment timing;
8Other protections include leakage, operation indication, automatic shutdown after fault alarm, etc.
High precision constant temperature and humidity laboratory | High precision constant temperature and humidity laboratoryUsage conditions:
1Installation site
The ground is flat and well ventilated
There is no strong vibration around the equipment
There is no strong electromagnetic field around the equipment
There are no flammable, explosive, corrosive substances or dust around the equipment
Leave appropriate space for use and maintenance around the equipment,
2Power supply conditions
Power requirements:AC380V±10% 50±0.5Hzthree-phase five-wire system
Pre installed power: total power+2.0KW
Require users to configure corresponding capacity air or power switches for the equipment at the installation site, and this switch must be independently used for the equipment (recommended power switch capacity:32A)
3Environmental conditions
Ambient temperature5℃~+30℃(24Average temperature within the hour ≤30℃)
ambient humidity:≤85%RH