Vibrio cholerae 01/0139 rapid detection kit

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The rapid detection kit for Vibrio cholerae 01/0139 is used for qualitative detection of Vibrio cholerae 01 and 0139 in fecal samples.

Product Details

Vibrio cholerae 01/0139 rapid detection kit

Common name:Vibrio cholerae assay kit(Colloidal gold method)

01 group test strips: 25 servings/box; 0139 group test strips: 25 servings/box

This product is used forRegarding fecesSample processing01 and 0139 Vibrio choleraeQualitative testing.

【 Principle of Inspection 】

This reagent kit uses colloidal gold immunochromatographic double antibody sandwich method to detect the presence of antibodies in specimensVibrio cholerae01, 0139 group antigens(Ag), Using colloidal gold as a marker and nitrocellulose membrane as a coating carrier. The nitrocellulose membrane is coated with two types of antibodies, one of which is antiVibrio cholerae01 or 0139 groupMonoclonal antibodies (Ab) are used to capture antigens in specimens; Another type is sheep anti mouse multi antibody, used for quality control of this test strip. Combined with other raw materials, assemble into test strips. During the detection process, the specimen is moved by the chromatographic action of the microporous membrane, and antigen antibody reactions occur during the movement. If the specimen containsVibrio cholerae01 or 0139 groupFirstly, it binds with colloidal gold labeled antibody (Au Ab) to form an Au Ab Ag complex. When the Au Ab Ag complex encounters the antibody coated on the membrane, it forms an Au Ab Ag Ab double antibody sandwich complex, presenting a visible red band.

If not present in the specimen01 or 0139 Vibrio choleraeThis area will not show color. Sheep anti mouse multi antibody zone, regardless of whether it is present in the sample or not01 or 0139 Vibrio choleraeAll display red, which serves as a marker for determining whether the test strip has failed or not.

Main components

Vibrio cholerae01 Group detection test stripThere are detection and quality control lines on the nitrocellulose membrane of the test strip.

Testing Line: Coating Resistance01 Vibrio cholerae groupMonoclonal antibody;

Quality control line: coated sheep anti mouse multi antibody;

Colloidal gold labeled antibody: anti01 Vibrio cholerae groupMonoclonal antibody.

Vibrio cholerae0139 group detection test stripsThere are detection and quality control lines on the nitrocellulose membrane of the test strip.

Testing Line: Coating ResistanceVibrio cholerae group 0139Monoclonal antibody;

Quality control line: coated sheep anti mouse multi antibody;

Colloidal gold labeled antibody: antiVibrio cholerae group 0139Monoclonal antibody.

Storage conditions and expiration date

Store at room temperature (4-30 ℃) and have a shelf life of 18 months.

Sample type:feces.

Sample processing:

  • Direct testing: Water samples are directly tested.2. Bacterial enrichment detection

Water sample stool:Inoculate 1-2 ml in 8-9 ml alkaline peptone water and incubate at 37 ℃ for 6-8 hours or overnight.

Solid fecal specimens: Inoculate 1-2g in 8-9ml alkaline peptone water, mix well, and incubate at 37 ℃ for 6-8 hours or overnight.

1. Before testing, the sample and test strip should be equilibrated to room temperature;

2. Place the test strip on a clean and dry workbench;

3. Take 2-3 drops of the sample using the pipette provided in this kit or a clean sample dispenser, and add them to the sample addition end of the test strip (white end of the test strip) to start timing;

4. Observation results for 2-15 minutes, reading results for 15 minutes or more is invalid.

Explanation of Inspection Results

Negative: A red line appears, indicating the presence of a red line at the position of the quality control line;

Positive: Two red lines appear, indicating the presence of red lines at both the detection and quality control lines;

invalid

  • No red lines appear at the positions of the detection line and quality control line.
  • The quality control line has not been released, only the detection line shows a red line.

Limitations of Inspection Methods

At present, the detection limit of this type of method is 105Individual bacteria per milliliter, therefore, it is necessary to conduct bacterial culture and detection to improve the detection rate. Because this reagent is effective for 104Bacteria with a concentration of less than/ml cannot be detected. When determining the result, only qualitative judgment can be made for positive or negative, and no quantitative judgment can be made.

Product Performance Indicators

  • The detection limit is 105Bacteria per milliliter.
  • Specificity: No cross reactivity was found with common pathogenic bacteria in the intestine. Such as Vibrio parahaemolyticus, Vibrio vulnificus, Vibrio parahaemolyticus, Vibrio alginolyticusEscherichia coliEscherichia coli0157:H7、 Salmonella typhi, Salmonella paratyphoid A, Salmonella enteritidis, Salmonella typhimurium, Shigella dysenteryYersinia enterocolitis, Proteus mirabilis, Bacillus subtilis, etc.
  • Stability: Room temperature (4-30 ℃), 18 months.

【 Precautions 】

  • Please read the instruction manual carefully before operation.
  • Judge the results under good lighting conditions.
  • Please useFresh water sample or stool specimenFresh cultured meat broth specimens should not be used, and old specimens should not be used.
  • Please use the pipette provided with this reagent kit or the self prepared clean sampler to drip the sample. If the test strip is directly immersed in the sample, it should not exceed 0.5cm.
  • The test strips should be stored in a dry environment, paying attention to moisture prevention.
  • Please use within the validity period of the test strip.
  • This test strip is only used for in vitro diagnosis.
  • Please pay attention to disinfecting contaminated areas and potentially infectious pollutants.
  • Low room temperature can appropriately prolong the reaction time.
  • Vibrio cholerae 01/0139 rapid detection kit