VOC testing climate chamber

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The VOC testing climate chamber can provide a standard testing environment (temperature, cleanliness, relative humidity, air displacement rate, etc.), and the VOC testing climate chamber tests the VOC/SVOC and moisture of various materials and products. Very suitable for specific analysis of heterogeneous prototype releases and detection of low trace components VOC detection climate chamber

Product Details

1VOC testing climate chamberpurpose:Detect the release of substances on the surface of materials at a constant temperature:

1. Modify the volatile matter and molecular material data before;

2. Compare product performance within an effective range (e.g. different colors/patterns);

3. Testing prototypes and low volatility materials;

4. During the formal certification testing period, test the product's * performance

IIVOC testing climate chamberstandard

GB18584-2001 "VOC detection for harmful substances in indoor decoration materials and wooden furniture in climate chambers"

GB18587-2001 "Indoor Decoration Materials - Carpets, Carpet Pads, and Carpet Adhesives - Release of Hazardous Substances *

ASTM D6007-02 "Standard Test Method for Formaldehyde Concentration in Gas Emissions from Wood Products in a Small Scale Environmental Chamber"

ENV717-1 "Environmental Chamber Method for Measuring Formaldehyde Emissions from Artificial Boards"

HJ 571-2010 "Technical Requirements for Environmental Labeling Products - Artificial Boards and Their Products"

HJ 566-2010 "Technical Requirements for Environmental Labeling Products - Wooden Toys"

Hazardous Substances in Soft Furniture Sofas (Draft for Comments)

Climate chamber for detecting volatile organic compounds in furniture (draft for comments)

LY/T1980-2011 "Volatile Organic Compounds (VOC) and Formaldehyde Release Detection Chamber

3、 Technical requirements for VOC detection climate chamber

1. Temperature: Adjustment range: (15-40) ℃

2. Adjustment accuracy: ± 0.5 ℃

3. Relative humidity: adjustable range: (40-70)% RH

4. Adjustment accuracy: ± 3% R.H

5. Air exchange rate: adjustable range: 0.2m3-2m3/h (air exchange rate 0.2-2 times/hour)

6. Adjustment accuracy: Ventilation rate ± 1%

7. Surface wind speed: The air horizontal velocity is maintained at (0.1-0.3) m/s, with an accuracy of ≤ 0.05m/s

8. VOC detection climate chamber size: (1 ± 0.02) m3

9. The test chamber is made of SUS304 mirror stainless steel plate welded as a whole. The welding area and surrounding areas are polished without deformation, and all edges

Make the corners into a certain curvature. Can be washed with water for cleaning, free from materials such as rubber and adhesives that interfere with testing, sealing materials, and other devices that come into contact with the air inside the testing chamber, and do not adsorb organic matter and formaldehyde. There is no condensation or condensation of water droplets in the test chamber.

10. Pressure: 10 ± 5Pa

11. Sealing: Static leakage rate: Under a relative pressure of 1KPa, the leakage rate is ≤ 1 × m3/min

12. Insulation box: Polyurethane insulation board with a thickness of 120mm, outer wall made of steel plate, white spray painted; Inner wall: 304 stainless steel plate

13. Background concentration: The concentration of particles in the gas is required to be less than 100 particles/m3 with a diameter greater than 0.5um

14. Formaldehyde background concentration<0.006mg/m3

TVOC background concentration<0.02mg/m ³

Single VOC background concentration<0.002mg/m3

15. Clean gas supply system:

(1) Compressed air: Volume: 2 m3/h;

(2) Drying machine: maximum air volume processing: 2m3/h;

(3) High efficiency particulate filter: maximum air flow rate: 2m3/h. After treatment, the concentration of pollutants in the gas is required to be less than 100 particles/m3 with a diameter greater than 0.5um.

4、 Introduction

This machine can provide a standard testing environment (temperature, cleanliness, relative humidity, air exchange rate, etc.) to test the VOC/SVOC and moisture of various materials and products. Very suitable for specific analysis of heterogeneous prototype releases and detection of low trace components