Two door cold and hot shock test chamberNature is looking for a manufacturer of high-quality reliability testing equipment, Kewen Testing Equipment Co., Ltd
Two door cold and hot shock test chamberTechnical parameters: Model: KW-TSE-80F/X/S KW-TSE-100F/X/S
KW-TSE-150F/X/S KW-TSE-250F/X/S
1. Inner box size: 500 * 450 * 450mm (width, height, depth)
Outer box size: 1380 * 1980 * 1370 (width, height, depth)
The size specifications can be customized according to customer requirements for cold and hot impact testing equipment solutions
2. Preheating temperature range: normal temperature to+200 ℃;
3. Pre cooling temperature range: normal temperature to -65 ℃; About 80 minutes (energy storage temperature range)
4. Preheating high temperature range: normal temperature~200 ℃; About 15 minutes; (Energy storage temperature range)
(Note: The cooling time refers to the performance when the low-temperature chamber is operated separately.)
5. Test method: Two box area;
6. Temperature shock range: -40 ℃ to+150 ℃
7. Temperature fluctuation: ± 0.3 ℃
8. Temperature deviation: ± 2.0 ℃
9. Temperature shock recovery time: ≤ 5 minutes (load; 3KG plastic)
10. Sample restriction: This experimental equipment prohibits the testing or storage of samples containing flammable, explosive, or volatile substances; Testing of Corrosive Substance SamplesOr storage; Testing or storage of biological samples; Testing or storage of strong electromagnetic emission source samples.
Introduction:
This box can simulate the instantaneous environmental changes between high and low temperatures. So as to determine whether the reliability and stability performance parameters of the product are qualified.
We will provide you with a basis for predicting and improving the quality and reliability of our products. Used for detecting electronics LED、 Automotive, rubber, plastic, aerospace technologyThe resistance of communication equipment and other products to repeated changes in temperature and heat.
Product Features:
*Upgraded in design with easy-to-use panel interface.
Adopting Japanese made * color touch LCD Chinese/English microcomputer temperature controller.
Vertical three box structure, automatic control of high and low temperature box circulation process, adjustable dwell and transition time, stainless steel inner tank, multi form recording, equipped withMultiple safety protection measures and devices.
*Protection alarm function: in case of short circuit, leakage, or overheating in the working room; Abnormal conditions such as compressor overpressure, overload, oil pressure, and water cut-offAt that time, the screen automatically displays the cause of the fault and the troubleshooting method. When the input voltage is found to be unstable, it has an emergency stop device and a cold and hot shockThe movement time of the structure is within 10 seconds. The temperature recovery time for cold and hot shock is within 5 minutes, which can comply with specifications such as MTL, IEC, JIS, GJB, etc.
Shenzhen manufactureroperationNotes:
1、 During operation, please do not open the box door unless necessary, as it may lead to the following adverse consequences.High temperature airflow rushes out of the boxVery dangerousThe inside of the box door still maintains high temperatureCausing burnsHigh temperature air may trigger a fire alarmGenerate misoperation
2、 Avoid shutting down and restarting the refrigeration unit within three minutes.
3、 Please note that the high and low temperature damp heat test chamber must be safely grounded to prevent static electricity induction.
4、 Circuit breakers, over temperature protectors, provide safety protection for local test products and operators, so please check regularly.
5、 If the heating test sample is placed inside the box, please use an external power supply for the sample power control, and do not directly use the high and low temperature wet heat test box power supply.
When placing high-temperature samples for low-temperature testing, attention should be paid to keeping the opening time of the box door as short as possible.
6、 Explosive, flammable, and highly corrosive substances are prohibited from being placed in the container for testing.
7、 The lighting should be turned off at all times except when necessary.
8、 On any section perpendicular to the prevailing wind direction, the sum of the cross-sectional areas of the test load should not exceed one-third of the working section at that location.
9、 When conducting high temperature tests, do not turn on the refrigeration unit when the temperature exceeds 55 ℃.
10、 Before doing low-temperature work, the studio should be dried dry and dried at 60 ℃ for 1 hour.