Rapid temperature change test chamberScope of application:
Rapid temperature change test chamberThis equipment is suitable for conducting environmental simulation reliability tests such as high and low temperature constant and gradient, wet and heat tests on various electrical and electronic products, as well as other products, components, and materials.
Rapid temperature change test chamberProduct Parameters:
| model | KW-KS-100 | KW-KS-150 | KW-KS-225 | KW-KS-408 | KW-KS-800 | KW-KS-1000 | ||
| Nominal internal volume (liters) | 100 | 150 | 225 | 408 | 800 | 1000 | ||
| performance | temperature range | T:-20~+150℃ | F:-40~+150℃ | S:-70~+150℃ | ||||
| humidity range | (25~98)%RH/(20~70)℃ | |||||||
| temperature fluctuation | ±0.5℃ | |||||||
| temperature deviation | ±2.0℃ | |||||||
| Humidity deviation | ±3.0%RH(>75%RH) ±5.0%RH(≤75%RH) | |||||||
| Range of temperature change rate | 5/10/15℃/min | |||||||
| internal dimensions (mm) | W | 500 | 600 | 600 | 800 | 1000 | 1000 | |
| H | 500 | 500 | 750 | 850 | 1000 | 1000 | ||
| D | 400 | 500 | 500 | 600 | 800 | 1000 | ||
| external dimensions (mm) | W | 930 | 1070 | 1070 | 1230 | 1430 | 1560 | |
| H | 1650 | 1670 | 1670 | 1880 | 2150 | 2150 | ||
| D | 950 | 1050 | 1050 | 1150 | 1350 | 1810 | ||
| Environmental temperature for use | +5~+35℃ | |||||||
| material | shell | Double sided galvanized steel sheet, surface spray treated | ||||||
| Internal body | Stainless steel plate SUS304 | |||||||
| adiabatic | Rigid polyurethane foam+glass fiber | |||||||
| regulator | heater | Nickel chromium alloy electric wire heater | ||||||
| humidifier | Stainless steel armored heater (heating and humidification method) | |||||||
| fan | centrifugal fan | |||||||
| refrigeration compressor | Piston compressor | |||||||
| refrigerant | R404A/R23 | |||||||
| control
manufacture
device | monitor | 7-inch TFT color LCD display | ||||||
| Operation mode | Program setting and fixed value method | |||||||
| Setting method | Chinese menu, touch screen input | |||||||
| Program capacity | Editable: 20 programs. Each program has a maximum of 50 steps and can have 20 loops (with a maximum of 99 loops per loop step); Fixed: 10 programs; The program can be connected. | |||||||
| input | thermocouple | |||||||
| communication function | RS-232 interface with local and remote communication capabilities | |||||||
| Curve recording function | Can save the set values, sampling values, and sampling time of the device | |||||||
| Supplementary functions | Fault alarm and cause, processing prompt function; Fault records; Overtemperature protection; Upper and lower temperature protection; Power outage protection; Test pause; Alarm output; Timer function (automatic start and automatic stop operation). | |||||||
| power supply | AC (380 ± 38) V (50 ± 0.5) Hz three-phase four wire+protective earth wire | |||||||
| Standard Configuration | One transparent electric heating film glass observation window, two test holes (customized according to actual size), one lighting lamp, four casters, one cable, and two shelves. | |||||||
| security configuration | Refrigeration system: compressor overpressure, compressor motor overheating, compressor current, cooling water supply under pressure Test box: Over temperature protection, temperature fuse, fan motor overheating Other: Main power phase sequence and phase loss protection, leakage protection, load short circuit protection | |||||||
Dongguan Kewen Experimental Equipment Co., Ltd. specializes in sales and productionHigh and low temperature test chamber、High and low temperature damp heat test chamber、constant temperature and humidity chamber、Rapid temperature change test chamber、Walk in high and low temperature test chamber, temperature and humidity vibrationThree comprehensive test chambers, electromagnetic vibration table, UV ultraviolet aging test chamberandIP Rain Level Test ChamberWaiting for equipment.
quality assurance
Provided by the supplier5 degrees per minute high and low temperature rapid temperature change test chamberThe product is brand new, and the overall system comes with a one-year quality guarantee, starting from the date of user acceptance.
Installation and debugging:
After signing the contract, the supplier shall provide the user with the equipment foundation diagram. After the equipment is in place, skilled personnel will be arranged to install and debug it on site (with the user assisting in the system installation).
Technical training:
Responsible for providing on-site training to users during the installation and commissioning process of the rapid temperature change test chamber equipment, and meeting the training requirements proposed by users based on actual situations until they can proficiently operate the equipment.
After-sales service
Within the warranty period, we will respond within 2 hours after receiving user fault information, troubleshoot within 48 hours, and provide free replacement of damaged parts and services. The after-sales service outside the warranty period is the same as the promise made during the warranty period, but the user bears a certain cost.
Kewen will visit users regularly or irregularly to provide experimental technology consultation, so as toUsers benefit the most from it.