| serial number | model | Basket size (mm) | Appearance (mm) | Power supply/total rate A |
| 01 | JY-JTS-80A | 400×400×500 | 1200×2500×1800 | 380V/8KW |
| 02 | JY-JTS-100A | 500×400×500 | 1400×2500×1800 | 380V/8.5KW |
| 03 | JY-JTS-150A | 500×500×600 | 1500×2600×1850 | 380V/9.5KW |
| 04 | JY-JTS-210A | 600×500×700 | 1700×2650×1900 | 380V/9.5KW |
| 05 | JY-JTS-260A | 630×550×750 | 1900×2700×1950 | 380V/10KW |
| 06 | JY-JTS-512A | 800×800×800 | 2500×3200×2100 | 380V/13W |
| 07 | JY-JTS-1000A | 1000×1000×1000 | 3000×3500×2100 | 380V/15KW |
| 08 | JY-JTS-1500A | 1500×1000×1000 | 4000×4200×2200 | 380V/20KW |
| temperature
| high-temperature chamber | RT~200℃ | ||
| Low-temperature chamber | RT~-70℃ | |||
| Impact temperature range | -Select 40, -50, -60 ℃~120, 150 ℃ as required | |||
| Temperature fluctuation ≤ ± 0.5 ℃ | Temperature uniformity ≤± 2 ℃ | Noise level (dB) ≤ 65 | ||
| Heating and cooling rate | warm up | 60min/ RT~200℃ | ||
| cool down | 60min/ 0~-70℃ | |||
| Experimental method | temperature conversion | The test piece basket is switched and transferred between high and low greenhouses for dynamic operation | ||
| Transfer method of test piece | pneumatic | |||
| Test piece transfer time | ≤5S | |||
| Temperature recovery time | ≤5min | |||
|
refrigeration | system | French fully enclosed Taikang compressor, cascade refrigeration (or valley wheel) | ||
| refrigerant | Cascade Level 2: R-23 (Fluorine free) | |||
| cooling | Air cooled (or water-cooled) | |||
| sensor | temperature | *Platinum resistor A (PT100) | ||
| humidity | *Platinum resistor A (PT100) | |||
| Box material | studio | #304 mirror stainless steel (1.2mm) | ||
| Outer box | Cold rolled steel plate electrostatic spraying (1.5mm) | |||
| insulation | Polyurethane+ultra-fine glass fiber (110mm below -50 ℃) | |||
| control | Japanese temperature control instrument, easy to operate, convenient to set, control accuracy of 0.01 ℃, programming beyond 1200 segments | |||
| heater | Fin type stainless steel heating system | |||
| evaporator | High efficiency heat exchanger | |||
| note | Non standard special requirements, can be customized | |||
other
▣ Operating environment temperature:+5~+35 ℃ (cooling environment temperature+5~+28 ℃ is preferred)
▣ Safety: Short circuit protection, overtemperature protection, compressor overpressure protection, compressor overload protection, fan overload protection, leakage protection, phase sequence and phase loss protection, operator safety protection, specimen protection
Attachment Standard Configuration
▣ Lighting: Philips 1
▣ High end wheels: 4