Turbidity and Solid Probe

NegotiableUpdate on 12/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Turbidity is caused by particles in the liquid. Turbidity meters use optical methods to measure the turbidity of media. For this purpose, a combination of transmitted/scattered light measurement is adopted, where the transmitter and receiver are placed opposite each other, and the other transmitter is arranged orthogonally (in a 90 #176; Corner). The absorption/scattering of transmitted light is recorded by the integrated evaluation electronic device in the connector, converted into a 4-20 mA signal, and output to a separate evaluation device. The measurement is carried out according to ISO 7027 (90 scattered light measurement).

Product Details

TC20,TC1000Turbidity and Solid Probe

浊度和固体探头

Turbidity is caused by particles in the liquid. Turbidity meters use optical methods to measure the turbidity of media. For this purpose, a combination of transmitted/scattered light measurement is adopted, where the transmitter and receiver are placed opposite each other, and the other transmitter is arranged orthogonally (at a 90 ° angle). The absorption/scattering of transmitted light is recorded by the integrated evaluation electronic device in the connector, converted into a 4-20 mA signal, and output to a separate evaluation device. The measurement is carried out according to ISO 7027 (90 ° scattered light measurement).


TC20、TC1000

The online turbidity measurement system is designed as a flow-through component.

  • Continuous turbidity measurement

  • Plastic version

  • Compact version - can be used without an additional transmitter

  • Compliant with ISO 7027 standard - Measurement of 90 ° scattered light and 180 ° transmitted light

  • Smooth measuring chamber - no particle settling

  • CLEANOSIL VMF Nano Window Coating - Long Maintenance Interval

  • Pre calibrate with Foremarazine (FNU unit)

  • Solid state calibration standards are included during delivery - simple equipment inspection and readjustment without the need for wet calibration

  • Customizable - easily select measurement range through DIP switch, no need for recalibration

  • Resolution - Starting from 0.001 FNU

probe type TC20、TC1000
power supply voltage 24V DC nominal (10... 30 V DC), electrically isolated from the measuring circuit
output signal 4... 20 mA (current absorber), such as B. used to connect MV5060A/B transmitters, KM3000 multi parameter measuring devices, or analog PLC inputs
Connect the load Approximately 0.5 W (note: if the 4... 20 mA measurement signal is provided by the connected measuring device, the connected load will increase by 24 V DC x 20 mA=0.48 W)
ambient temperature 0... 45 degrees Celsius
medium temperature 0... 60 degrees Celsius
pressure 6 bar/20 ℃; 1 bar/60 ℃
sealing element EPDM rubber
Sensor Window Borosilicate glass, Cleanosil coating
signal Status LED
measurement range TC20:0.01... 1/2/5/10/20 FNU

TC1000: Optional 0.1... 50/100/200/500 FNU or 100... 1000 FAU
measurement accuracy ± 5% of the current measurement value and ± 1% of the corresponding measurement range final value
solve 0.001 … 0.2 FNU, Depending on the measurement range
process connection DN25, Pipe fittings with adhesive sockets
CE mark Compliant with EMC Directive (89/336/EEC) and Low Voltage Directive (2006/95/EC)
Color/Pollution Compensation Integrated into TC20
application Well water monitoring, raw water monitoring, cooling water monitoring, filter backwash monitoring, sewage treatment plant outlet control, centrifuge monitoring, filter monitoring, phase separation
Order Description TC20、TC1000