E-mail
977889474@qq.com
- Phone
-
Address
No. 159-9 Jinhu West Road, Jinhu Economic Development Zone, Jiangsu Province
Jiangsu Suyi Group Co., Ltd
977889474@qq.com
No. 159-9 Jinhu West Road, Jinhu Economic Development Zone, Jiangsu Province
Toxic gas detector (carbon monoxide)
Working principle of toxic gas detector (carbon monoxide)
During operation, sensors sample gases and transmit real-time data to high-speed processing chips for real-time analysis and processing. Then it is transmitted in real-time to the main chip, which processes it into switch signals, analog signals, or digital signals for users to choose from. This instrument can still continuously detect gases in unmanned hazardous environments and can be widely used in places such as petroleum, chemical, smelting, pharmaceutical, oil fields, oil depots, storage, environmental protection, fire protection, municipal and other places where there may be dangerous explosive gases. It is a monitoring instrument for personal safety and equipment safety.
Technical specifications of toxic gas detector (carbon monoxide):
range0-1000ppm
Indication error: ≤±3%F.S
Small reading:1ppm
Response time: ≤30second
Sensor lifespan:24Months (imported sensors)
Detection principle: electrochemistry
Detection method: diffusion type
Working mode: Continuous work
Working voltage:24V DC
Explosion proof electrical interface:M20*1.5Internal thread (optional)G3/4 G1/2Conversion connector)
Structural material: die cast aluminum
Explosion proof mark:EXd II CT6
Operating Temperature-40~70℃
Working humidity: ≤90%RH
size180*190*95mm
weight 1580g
transmission distance ≤1500mm
connecting cable Three core shielded wire ≥RVVP 3*1.0mm2(National standard line)
Installation method Fixed bracket, pipe installation, wall mounting
output signal4-20mACorresponding to current signal0full scale 2Group switch outputRS485Digital signal output
Fault Alarm The system can display gas detection alarm wiring short circuit, open circuit fault, sensor fault, and system power supply fault
Measurable toxic and harmful gases
serial number |
name |
molecular formula |
measurement range(ppm) |
response time(second) |
recovery time(second) T10 |
|
T50 |
T90 |
|||||
D001 |
oxygen |
O2 |
0-25%v/v |
<10 |
<25 |
<20 |
D002 |
carbon monoxide |
CO |
0-100/200/500/1000 |
<15 |
<25 |
<45 |
D003 |
hydrogen sulfide |
H2S |
0-20/50/100 |
<10 |
<30 |
<40 |
D004 |
hydrogen |
H2 |
0-1000 |
<20 |
<50 |
<60 |
D005 |
ammonia |
NH3 |
0-100/500/1000 |
<40 |
<90 |
<120 |
D006 |
chlorine gas |
CL2 |
0-10/20 |
<30 |
<50 |
<60 |
D007 |
nitric oxide |
NO |
0-100/500 |
<10 |
<30 |
<90 |
D008 |
nitrogen dioxide |
NO2 |
0-20/50 |
<40 |
<90 |
<90 |
D009 |
sulfur dioxide |
SO2 |
0-20/100 |
<15 |
<40 |
<90 |
D010 |
vinyl chloride |
C2H3CL |
0-20/100 |
<35 |
<90 |
<180 |
D011 |
acrylonitrile |
C3H3N |
0-100/200 |
<80 |
<160 |
<180 |
D012 |
methanol |
CH3OH |
0-100/200/500 |
<20 |
<80 |
<150 |
D013 |
formaldehyde |
CH2O |
0-50/100/200/500 |
<25 |
<70 |
<150 |
D014 |
Ethylene Oxide |
C2H4O |
0-20/100/200 |
<50 |
<120 |
<180 |
D015 |
hydrogen chloride |
HCL |
0-30/50 |
<50 |
<80 |
<120 |
D016 |
hydrogen fluoride |
HF |
0-10 |
<30 |
<90 |
<120 |
D017 |
hydrogen cyanide |
HCN |
0-30 |
<40 |
<90 |
<120 |
D018 |
sulfur trioxide |
SO3 |
0-100 |
<30 |
<90 |
<120 |
D019 |
phosgene |
COCL2 |
0-1 |
<30 |
<90 |
<120 |
D021 |
acrylonitrile |
C3H3N |
0-100 |
<30 |
<90 |
<120 |
D022 |
carbon disulfide |
CS2 |
0-100 |
<30 |
<90 |
<120 |
D023 |
Arsine |
AsH3 |
0-10 |
<30 |
<90 |
<120 |
D025 |
Bromine gas |
Br2 |
0-20 |
<30 |
<90 |
<120 |
D027 |
Ethylene Oxide |
C2H4O |
0-100 |
<30 |
<90 |
<120 |
D028 |
Tetrahydrothiophene |
C4H8S |
0-25 |
<20 |
<60 |
<120 |