KB-6portableparticle counter
Overview:
Offline detection:
Take samples directly from the sample bottle for testing; Connect the pressure bottle sampler (pressurization device) for sampling and testing.
Online detection:
It can be installed on various hydraulic transmission, lubrication, oil filter, cleaning machine, testing bench and other systems to achieve online detection of system oil contamination (cleanliness).
Tianjin Logan KB-6 portable particle counter-Performance characteristics:
1. Adopting the principle of light blocking method, it has the advantages of fast detection speed, strong anti-interference ability, high accuracy, and good repeatability.
2. High precision sensors ensure high resolution and accuracy.
3. Precision measurement sampling system, achieving constant sampling speed and precise control of sampling volume.
4. 7-inch color Android system LCD screen display, touch screen operation.
5. Built in NAS1638、ISO4406-1999、ISO4406-2017、ISO4406-1987、GB/T 14039-2002、GJB420B-2006、GJB420B-2015、GJB420A-1996、AS4059E Differential, AS4059E cumulative, AS4059F differential, AS4059F cumulative ROCT17216-1971、ROCT17216-2001、JTS-005、SAE749D-1963、AETF1A、NAS1638/ISO4406 According to the particle pollution level standards, the required standards can be built in according to user requirements.
6. Built in "custom" standard, particle size can be set arbitrarily at intervals of 0.1 μ m, with 990 particle sizes available for easy particle size analysis.
7. Built in "full channel" standard, which can simultaneously detect the particle size of all detection standards. After the detection is completed, multiple detection standards can be selected arbitrarily, and the detection report can be displayed and printed according to the selected multiple detection standards.
8. Built in calibration function, can be calibrated according to ISO4402, GB/T18854-2002 (ISO11171:1999), ROCT and other standards.
9. It can simultaneously store three calibration curves (ACFTD calibration curve, ISOMTD calibration curve, GOST calibration curve) and can be easily switched to reduce conversion errors.
10. Testing data storage function, 6GB of memory, unlimited storage of data, convenient for archiving, retrieval, and analysis of testing data. And equipped with a USB interface, data can be stored on a USB drive.
11. Built in printer that can directly print out inspection reports.
12. Built in multiple input methods, capable of handwritten input, enabling Chinese annotation of detection reports.
13. The output interface includes RS232 or RS485, dual USB interface, and LAN interface, with optional Bluetooth interface; It can be connected to USB flash drives, optional A4 format printers, optional barcode scanners, and other peripherals. It can also be equipped with computer data acquisition software to enable external computers to annotate, store, query, and generate PDF detection reports for detection data.
14. Equipped with built-in or external lithium batteries, outdoor testing can be easily completed.
15. The instrument can achieve continuous automatic online detection, provide real-time particle number and grade, and can set the detection interval time arbitrarily.
16. You can choose to install a water activity sensor to detect water activity (aw value), water content (ppm), and oil temperature in the oil.
17. It has a flushing function and the flushing volume can be set arbitrarily.
18. Adopting high-strength injection molded casing, the structure is compact, lightweight, and easy to carry.
19. Adopting high-precision aluminum brushed panels, simple and beautiful, durable and long-lasting.
Tianjin Logan KB-6 portable particle counter-Technical Indicators:
1. Light source: semiconductor laser
2. Detection range: 1 μ m to 600 μ m (depending on the selected sensor)
3. Sensitivity: 1 μ m (ISO4402) or 4 μ m (c) (GB/T18854, ISO11171)
4. Detection channels: 128, with an interval of 0.1 μ m, allowing for arbitrary setting of particle size
5. Detection method: offline or online
6. Sampling volume: 10-1000ml/time
7. Relative error of sampling volume: better than ± 3%
8. Testing time: 30s/time
9. Detection speed: 5-60mL/min (default 20mL/min)
10. Cleaning speed: 5-60mL/min (default 20mL/min)
11. Rinse volume: can be set between 0ml and 90ml (with an interval of 1ml)
12. Overlap error limit: 12000~40000 particles/mL (depending on the selected sensor)
13. Particle count repeatability: RSD<2%
14. Relative error of particle counting: ± 10%
15. Offline viscosity detection: ≤ 100cSt (optional pneumatic bottle sampler viscosity can reach 400cSt)
16. Online pressure detection: 0.1-0.6Mpa (with optional pressure reducing device, the maximum pressure can reach 40MPa)
17. Online detection interval time: can be set between 1 second and 10 hours, 59 minutes and 59 seconds
18. Test sample temperature: 0 ℃~80 ℃
19. Working temperature: -20 ℃~60 ℃
20. Storage temperature: -30 ℃~80 ℃
21. Power supply: 100V~265VAC or 24VDC or lithium battery
22. Dimensions: 410 × 330 × 170mm
23. Net weight of the instrument: 7kg