The Logan Kexing KT-2A oil contamination level detector is equipped with Tianjin Logan Kexing's advanced photoresist principle laser sensor, which can detect the size and quantity of solid particles in transparent liquids, and determine the pollution level according to different detection standards, meeting the detection of liquid particle contamination levels in various industries. The nKT-2A particle counter is equipped with calibration methods such as ACFTD, ISOMTD, GOST, latex balls, etc., and has more than 20 oil contamination detection standards built-in. The nKT-2A particle counter is equipped with a positive/negative pressure chamber device to ensure that it does not
The Logan Kexing KT-2A oil contamination level detector is developed based on the principle of light resistance (shading method), and is used to detect the size and quantity of solid particles in liquids. It can be widely used in aviation, aerospace, aviation, heavy machinery, power, petroleum, chemical, transportation, ports, metallurgy, machinery, automobile manufacturing and other fields. It is used to detect the solid particle contamination level of hydraulic oil, lubricating oil, shale oil, transformer oil (insulation oil), turbine oil (turbine oil), gear oil, engine oil, aviation kerosene, water-based hydraulic oil and other oils, as well as to sample and detect insoluble particles in water-based samples, chemical reagents, organic liquids, polymer solutions and other oils.
Oil contamination level detectorThe 'invisible guardian' that safeguards the health of devices
In modern industrial production, oil serves as an important lubricant and cooling medium for mechanical equipment, and its cleanliness directly affects the operational efficiency and lifespan of the equipment. Therefore, the detection of oil contamination has become crucial. The oil contamination level detector, as a professional tool in this field, is becoming an "invisible guardian" in industrial production due to its high precision and efficiency.
1、 Working principle
The oil contamination level detector mainly detects particulate pollutants in oil through optical, electronic, or physical methods. Its working principle usually includes steps such as sampling, dilution, filtration, detection, and analysis. Firstly, take a sample from the oil to be tested, then dilute and filter it to remove larger particles and impurities. Next, the processed oil sample is sent to the detection area of the detector, where signals of particulate pollutants are captured by optical or electronic sensors and converted into readable values or images. Finally, based on the test results, evaluate and analyze the degree of contamination of the oil.
2、 Characteristics
High precision: Advanced sensors and algorithms are used to accurately detect various sizes and types of particulate pollutants in oil, including metals, non metals, fibers, etc. At the same time, its measurement range is wide, which can meet the detection needs of different industries and equipment.
High efficiency: Compared to traditional oil contamination detection methods, oil contamination meters have higher detection efficiency. It can complete the detection of a large number of samples in a short period of time, greatly improving work efficiency. In addition, the detector also has automatic cleaning and calibration functions, reducing manual operation and maintenance costs.
Easy to operate: The oil contamination level detector usually uses a touch screen or computer interface for operation, with a simple and clear interface and convenient operation. Users only need to follow the prompts to complete the detection process, without the need for complex training and learning.
Intelligence: Modern oil pollution meters usually have intelligent functions, such as automatic recording, data analysis, report generation, etc. Users can manage and analyze detection data through software, timely discover and solve equipment failure problems, and improve the reliability and stability of equipment.
3、 Application Fields
The oil pollution level detector is widely used in equipment maintenance and fault diagnosis in industries such as automotive, aviation, power, metallurgy, and chemical. In the automotive industry, it can help maintenance personnel quickly determine the degree of oil contamination in components such as engines and transmissions, replace and clean the oil in a timely manner, and ensure the normal operation of the vehicle. In the aviation field, oil contamination meters can be used for oil detection of aircraft engines and hydraulic systems to ensure flight safety. In the power and metallurgical industries, it can help companies monitor the oil pollution level of transformers, cooling systems, and transmission systems, detect equipment failures in a timely manner, and take measures for maintenance.
4、 Summary and Prospect
The oil contamination level detector, as an important detection tool in industrial production, plays a crucial role in ensuring equipment health and improving production efficiency.

Oil contamination level detector
Logan Tech KT-2AOil contamination level detectorPerformance parameters:
1. Light source: semiconductor laser
2. Detection range: 0.8 μ m to 600 μ m (depending on the selected sensor)
3. Particle size range: 1-100 μ m (ISO4402) or 4-70 μ m (c) (GB/T18854, ISO11171)
4. Sensitivity: 0.8 μ m (ISO4402) or 3 μ m (c) (GB/T18854, ISO11171)
5. Detection channels: 16, with particle size adjustable at intervals of 0.01 μ m
6. Sampling volume: 0.2-100ml, with an interval of 0.1ml
7. Relative error of sampling volume: better than ± 0.5%
8. Detection speed: 5-80mL/min (default 15mL/min)
9. Cleaning speed: 5-80mL/min (default 15mL/min)
10. Resolution: better than 10% (GBT18854-2002, ISO11171)
11. Overlap error limit: 12000~40000 particles/mL (depending on the selected sensor)
12. Particle count repeatability: RSD<2%
13. Relative error of particle counting: ± 5%
14. Maximum negative pressure of the pressure chamber: -0.08MPa

Oil contamination level detector