| Main technical parameters and performance | temperature range | -70~ ﹢200℃ |
| temperature fluctuation | ±0.5℃(When the temperature is constant) | |
| Temperature Uniformity | ±2℃(When the temperature is constant) | |
| Maximum impact range | -40 ~+150℃ | |
| Switching time between high and low temperatures | ≤10S | |
| High and low temperature recovery time | ≤5min | |
| Heating rate of high-temperature chamber | From room temperature to+200℃≤45min(When running alone) | |
| Cooling rate of low-temperature chamber | From room temperature to minus zero-70℃≤75min(When running alone) | |
| studio dimensions | 700×1000×500mm(Deep)×wide×High) | |
| cooling system | water cooling | |
| Experimental box specifications | Three boxes of test objects stationary | |
| average wind speed | ≤1.7 m/s | |
| total power | about 43KW | |
| power supply | 380V±38V 50Hz±1Hz | |
| other | Program editing, running, and recording can be stored on a computer for analysis and printingIndia | |
| refrigerant | eco-friendly R404ATheR23 | |
| meet the standard | GB2423.22-89Na,GJB150.5-86,MIL-STD-810-503.2 |
1refrigeration system
1The refrigeration compressor adopts two branded semi enclosed compressors to form a cascade refrigeration system. This fully enclosed compressor has the characteristics of large refrigeration capacity, low energy consumption, low noise, and high conversion rate; The refrigeration system also adoptsPLC control, energy regulation cold end output method, surpasses the traditional cold and hot balance control method, meets requirements through PID regulation, greatly reduces user usage costs, extends equipment service life, and improves equipment economy; Beneficial for reducing equipment noise and improving the working environment of the laboratory.
2Cooling method: water cooling;
3Condensation evaporator: using a heat exchanger, which has a small volume, high temperature and pressure resistance, and a high heat transfer coefficient;
4Refrigerant: Environmental ProtectionR404A、R23;
Refrigeration auxiliary components: In order to ensure the safe and reliable operation of the refrigeration system, the main refrigeration accessories such as expansion valves, solenoid valves, filters, pressure controllers, globe valves, oil pressure controllers, oil separators, etc. are all branded products.