NegotiableUpdate on 12/19
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Three channel laboratory pH/conductivity/dissolved oxygen analyzer standard configuration $r $n • PE-30 six in one water quality detector $r $n • E201-BNC-10K triple composite pH electrode $r $n • pH standard buffer reagent (pH 4.00, 6.86, 9.18) $r $n • CON-1-10K platinum conductivity electrode $r $n • conductivity standard solution (84 S/cm, 1413 S/cm, 12.88mS/cm) $r $n • DO100 dissolved oxygen electrode $r $n • electrolyte $r $n • electrode membrane cap
Product Details

Three channel laboratory pH/conductivity/dissolved oxygen analyzermeasurement parameters

• pH, Conductivity, TDS, salinity, dissolved oxygen, temperature


Functional characteristics of pH unit

Circular pH sensitive membrane, fully sensing the pH value of the liquid.

Calibration from 1 to 3 points, automatically identifying USA or NIST standard buffer solutions.

Automatic temperature compensation ensures accurate measurement across the entire range.

The settings menu allows customization of pH buffer standards, number of calibration points, temperature units, etc.


Functional characteristics of EC unit

Platinum conductive electrode provides accurate and reliable measurement results.

Calibration from 1 to 3 points, automatic identification of conductivity standard solution.

Automatic temperature compensation, correcting conductivity measurements and converting readings to reference temperature.

The settings menu allows customization of the number of calibration points, TDS coefficient, temperature units, etc.


Functional characteristics of DO unit

Equipped with a polarographic electrode.

1 or 2-point calibration, using air saturated water or zero oxygen solution.

Salinity and atmospheric pressure compensation eliminate measurement errors.

Automatic temperature compensation ensures accurate measurement across the entire range.

The settings menu allows customization of the number of calibration points, concentration units, temperature units, etc.


General functional features

Automatic endpoint locking, retaining stable readings for easy browsing and recording.

Automatic power shutdown effectively saves battery energy consumption.

Reset function, automatically restore the instrument to factory default settings.

Three channel laboratory pH/conductivity/dissolved oxygen analyzer

parameter

PHunit

ECunit

DO Unit

PH

measurement range

-1.00~15.00pH

display resolution

0.01pH

measurement accuracy

±0.01pH

calibration point

From 1 to 3 o'clock

PH buffering option

USA (pH4.01/7.00/10.01), NIST (pH4.01/6.86/9.18)

conductivity

measurement range

0~20.00, 200.0, 2000µS/cm, 20.00mS/cm

display resolution

0.01, 0.1, 1

measurement accuracy

±1% F.S.

calibration point

1-3 o'clock

Use calibration solution

84µS, 1413µS, 12.88mS,

TDS

measurement range

0~10.00, 100.0, 1000ppm, 10.00, 200.0ppt

display resolution

0.01, 0.1, 1

measurement accuracy

±1% F.S.

TDS conversion coefficient

0.1 to 1.0 (default 0.5)

salinity

measurement range

0~10.00ppt

display resolution

0.01ppt

measurement accuracy

±1% F.S.

Dissolved oxygen concentration

measurement range

0.0~20.0mg/L

display resolution

0.1mg/L

measurement accuracy

±0.5mg/L

%saturation

measurement range

0.0~200.0%

display resolution

0.1%

measurement accuracy

±2%

dissolved oxygen

calibration point

1or2point

Range of atmospheric pressure compensation

450 ~ 850 mmHg,manual

Salinity compensation range

0~35g/L,manual

temperature

measurement range

060°C/32~140°F

display resolution

0.1°C/0.1°F

measurement accuracy

±1°C/±1.8°F

Offset calibration

1 point, measured value ± 10 ° C

other

Temperature compensation range

060°C/32~140° F, automatic

0~100°C/32~212°F, automatic

0~40°C/32~104°F,automatic

Temperature compensation coefficient

2%/°C, Linear

reference temperature

25°C

cell constant

K=1

Data locking

handMoving or automatic

Auto Shutdown

40After minutes of no button operation

Power type

AC220V/50Hz

Overall dimensions

210(long)×188(wide)×60(tall)mm

weight

1.5kg