Thermal resistance PT100

NegotiableUpdate on 05/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

When there is a step change in temperature, the time required for the output of the thermal resistor PT100 to change to 50% of the step change is called the thermal response time, represented by T0.5.

Product Details

Thermal resistance PT100Shanghai Instrument Third FactoryTemperature sensing element at 100 ℃Resistance value at time (R100)And it's at 0Resistance value R at time0The ratio: (R100/ R0

Division number Pt100: A-level R0= 100±0.06Ω

B-grade R0= 100±0.12Ω

Index number Cu50: R0=50 ± 0.05 Ω

Thermal response time

When there is a step change in temperature, the time required for the output of the thermal resistor to change to 50% of the step change is called the thermal response time, which is T0.5express.

Thermistor nominal pressure

Thermal resistance PT100Shanghai Instrument Third FactoryGenerally refers to the external pressure (static pressure) that the protective tube can withstand at the working temperatureDurable and unbreakable. The allowable nominal pressure is not only related to the material, diameter, and wall thickness of the protective tube, but also to its structural form, installation method, placement depth, as well as the flow rate and type of the measured medium.

Small insertion depth of thermistor

min=yuan+15D

Min - Small available insertion depth

yuan- Length of temperature sensing element

D - outer diameter of protective tube

natural influence

When measuring a current of 5mA through a thermal resistor, the measured resistance increment converted to temperature value should not exceed 0.30 ℃.

insulation resistance

The test voltage for insulation resistance at room temperature can be taken as 10 DC100V at any value, with an ambient temperature of 15Within the temperature range of 35 ℃, the relative humidity should not exceed 80%. The insulation resistance value at room temperature should not be less than 100MΩ

working principle

Industrial thermal resistors are divided into two categories: platinum thermal resistors and copper thermal resistors.

Thermistor measures temperature by utilizing the characteristic that a substance's own resistance changes with temperature. The heated part (temperature sensing element) of a thermal resistor is uniformly wound around a skeleton made of insulating material with thin metal wires. When there is a temperature gradient in the measured medium, the measured temperature is the average temperature in the dielectric layer within the range of the temperature sensing element.

Prefabricated thermistor mainly consists of a junction box, protective tube, terminal block, insulation sleeve, and temperature sensing element, and is equipped with various installation and fixing devices.

The temperature sensing element of WZP type platinum resistor is a platinum wire winding, and dual platinum resistors are mainly used in situations where two sets of display, recording, and regulating instruments are required to simultaneously detect the temperature at the same location. The temperature sensing element of WZC type copper resistor is a copper wire winding.

Model

Unified design type thermistor

Non fixed device type thermistor

Thermal resistance category

Product model

graduation mark

Measurement range ℃

Protective tube material

Diameter dmm

Thermal response time T0.5S

Single platinum thermistor

WZP-120

Pt100

(BA1、BA2) ﹡

-200~420

stainless steel

1cr18Ni9Ti

stainless steel

0cr18Ni12M0Ti

F16

≤90

WZP-121

F12

≤30

WZP-130

F16

≤90

WZP-131

F12

≤30

Double platinum thermistor

WZP2-120

F16

≤90

WZP2-121

F12

≤45

WZP2-130

F16

≤90

WZP2-131

F12

≤45

Copper thermistor

WZC-120

Cu50

(G)﹡

-50~100

Brass H62 stainless steel 1cr18Ni9Ti

F12

≤120

WZC-130

Note: 1. Mark the "*" division mark for special specification ordering

2. Models 120 and 121 are splash proof junction boxes, while models 130 and 131 are waterproof junction boxes.

Thermal resistance length specification

F16

F12

Total length LMM

Depth IMM

Total length LMM

Depth IMM

300

350

400

450

550

650

900

1150

1400

1650

2150

150

200

250

300

400

500

750

1000

1250

1500

2000

225

250

300

350

450

550

650

900

1150

75

100

150

200

300

400

500

750

1000