- Phone
-
Address
606, Block B, Chengyuan Building, Building 3, Jiancai Chengzhong Road, Haidian District, Beijing
Ansaisi (Beijing) Technology Co., Ltd
606, Block B, Chengyuan Building, Building 3, Jiancai Chengzhong Road, Haidian District, Beijing
Product Introduction:
NamePhase12BThermal analyzer
Model:Phase 12B
Brand:AnalysisTech
Origin: United States
Application:Phase 12BisPhase 12thermal analysisInstrument seriesThe subset. standardPhase 12Thermal analyzerandPhase 12BThe main difference between them isPhase 12BThere is an internal100APower booster.
Product Introduction:
Phase 12BisPhase 12thermal analysisInstrument seriesThe subset. standardPhase 12Thermal analyzerandPhase 12BThe main difference between them isPhase 12BThere is an internal100APower booster.Phase 12BThe high current port on top exists in a single100AWithin the range, the maximum voltage is20V. Stage12BThe remaining parts and stages of the standard 12 The standards are the same, butHere it isfind
Phase 12BIntended to collaborate withKeysight GPIBPaired with controlled power supply, the latter is included inPhase 12BIn the discount attachment package.
Keysight GPIBThe availability of controlled power supplyCommunication input optionsincluding:
· 230VAC, single-phase(190 - 240VAC,50/60HZ)
· 208VAC, three-phase(190 - 240VAC,50/60HZ)
· 400VAC, three-phase(380 - 415VAC,50/60HZ)
Phase 12and12BElectrical specifications:
All standardsThermal analyzerAll provided 2、10 or 20 The maximum heating current capacity of Ampere is determined by placing it after the model number“-2”、“-10”or“-20”To express.Phase12BIt's standardPhase12Thermal analyzerThe high current version is only available 100 The maximum heating current capacity in amperes. The following specifications indicate 2、10、20 and 100 Ampere unitWhere there are differences between them;Otherwise, the specifications apply to all versions. The requirements for AC power supply are assumed to be120 VAC,50/60HzUnless otherwise specified at the time of ordering. For higher power testing,Power Booster You can use electricityExpanding streaming capacity to1000AAnd expand the heating power output to4500W.
|
|
Heating current,high rangeprecision±0.2%Reading addition,
2ASystem:±0.2mA; 10ASystem:±1mA; 20ASystem:±2mA; 100ASystem(Phase12B):±10 mA
Heating current, low range accuracy:±0.2%Reading addition,
2ASystem:±0.02 mA; 10ASystem:±0.1mA; 20ASystem:±0.2mA; 100ASystem(Phase 12B):±10 mA
Heating voltage measurement accuracy:
Phase12: Reading of±0.2% ±Full range0.025%(Usually)50V)
Phase12B: Reading of±0.2% ±5mV(20VFull scale)
Thermocouple measurement accuracy(TType standard):±0.15°C(Typical value)±0.3°C(Maximum value)
Accuracy and precision of junction temperature measurement:Tj Digital resolution:±0.007°C(Typical value) Tj Display accuracy:±0.1°C
Tj Absolute accuracy:±0.2to±0.5°C(Typical values (see accuracy analysis)))
Delay in junction temperature measurement:minimum1 microsecond AC power supply:120VACThe3AThe50/60Hz(Unless otherwise specified, as standard)
Maximum heating power supply voltage:50 Volts (other voltages available),20VwithPhase 12B
Maximum heating power supply current:2AThe10AThe20Aor100A(Phase 12B), nonepower amplifier
Tj Induction current selection:1mAThe5mAThe10mAThe20mAThe50mAAnd adjustable0.1 - 50mA
Data collection sampling rate:1 MHz
Phase 12/12BChassis size:17 inchx 17inchx 7.5Inches high,25 pound
requiredPCHardware:
Create a system:Windows 7orWindows 10Professional version,32/64Position, English version (including)
processor3 GHzSingle coreCPUOr higher,2 GB RAMmemory
Graphics: Any installation
WindowsThe main graphics card of the driver program Others:500 GBHard drive, at least 4 aUSBport