Teclis foam analyzer (intelligent modular design)Specialized research on the performance of foam can not only obtain solutions for scientific foam analysis, but also scientifically optimize solutions for foam formation or foam prevention products and processes.

FOAMSCAN foam analyzer has intelligent modular design, which can describe:
• Gas is injected into liquid through porous glass filter ① to produce foam. Mass flow controller ② precisely controls gas flow rate.
• Stir the liquid ③ at a controlled speed to produce foam
• foam generated externally ④.
The measurement parameters are 100% controlled by software ⑤ to ensure the reproducibility of the experiment.
Once the measurement starts, the foam rises into the cylindrical glass measuring tube ⑥. Calculate foam volume in real time through image analysis ⑦. Real time calculation of liquid volume and foam liquid fraction through conductance. The second camera ⑧ captures the foam structure image and analyzes the bubble size and distribution.
The automatic cleaning system saves a lot of time. The cleaning sequence can be set before or after measurement to avoid removing the tube for cleaning.
The automatic cleaning system is connected to two liquids other than water. After the measurement, the built-in waste pump ⑩ discharges the sample from the foam column. The second pump is used to provide water and/or cleaning fluid at the top of the foam column ≮ to clean the foam column between measurements.
The Pt100 sensor measures the temperature inside the liquid sample. If the constant temperature water bath is connected to FOAMSCAN ™ Above, FOAMSCAN should be equipped ™ The quick disconnect connector provided in the attachment should be connected to one side of the instrument.
Teclis foam analyzer (intelligent modular design)Placed inside the protective cover ⑫, it can avoid light interference and protect the instrument from dust.

1. Gas jet foaming 2 Mass flow controller 3 Mechanical stirring foaming 4 Externally generated foam 5 Electronic indicators and software
6. Cylindrical glass measuring tube7. Main CCD camera 8 Second CCD camera 9 Fluid connection 10 Pump 11 Automatic cleaning system 12. Protective cover