Sulfide acidification blowing device

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The HCA-310 water sulfide acidification blowing device is a sulfide acidification blowing device that meets the pre-treatment standards for water sulfide samples. The sulfides in the water sample are acidified, and the generated hydrogen sulfide enters the absorption bottle/tube with the carrier gas (nitrogen) and is absorbed by the absorption solution (zinc acetate sodium acetate solution or 2% * solution). The corresponding analysis method (GB/T16489-1996 "Determination of Sulfides in Water Quality - Methylene Blue spectrophotometric method", HJ/T600-200 "Determination of Sulfides in Water Quality - Iodometric method") is selected to analyze the sulfur ions absorbed in the absorption bottle/tube

Product Details

Product Overview:



HCA-310Sulfide acidification blowing deviceisMeet the pre-treatment standards for sulfide samples in water qualitySulfide acidification blowing device.

  The sulfides in the water sample are acidified, and the generated hydrogen sulfide enters the absorption bottle/tube with the carrier gas (nitrogen) and is absorbed by the absorption solution (zinc acetate sodium acetate solution or 2% * solution). The corresponding analysis method (GB/T16489-1996 "Determination of Sulfides in Water Quality - Methylene Blue spectrophotometric method", HJ/T600-200 "Determination of Sulfides in Water Quality - Iodometric method") is selected to analyze and determine the sulfur ions absorbed in the absorption bottle/tube. Suitable for the determination of sulfides in surface water, groundwater, domestic sewage, and industrial wastewater.

  Using a constant temperature water bath for heating, PID precise temperature control, and a fast automatic constant temperature control system to shorten analysis time. The hydrochloric acid inlet, gas inlet, and outlet are integrated and independent of each other, and their operations do not interfere with each other. Automatic liquid level detection and replenishment of water bath system. Large capacity reaction bottle, independent six unit module, capable of processing six samples simultaneously. The nitrogen flow rate and blowing time of each sample are independently controlled. The large LCD touch screen display control is clear at a glance.


硫化物酸化吹气装置



Execution standard:



  • HJ/T 60-2000 "Determination of Sulfides in Water Quality - Iodometric Method"

    GB/T 16489-1996 "Determination of Sulfides in Water Quality - Methylene Blue spectrophotometric method"


Performance characteristics:



Large capacity reaction bottle, independent six unit module, capable of processing six samples simultaneously.

Constant temperature water bath heating, PID precise temperature control, fast automatic constant temperature control system.

The hydrochloric acid inlet, gas inlet, and outlet are integrated and independent of each other, and their operations do not interfere with each other.

Automatic liquid level detection and replenishment of water bath system.

Independent control of nitrogen flow rate and blowing time, automatically shutting off the gas source when the set time is reached, saving nitrogen without the need for personnel to shut it off.

Adopting a 7-inch color touch screen microcomputer human-machine interaction system, various instrument parameters can be independently set.

硫化物酸化吹气装置


Technical specifications:



Sample quantity: 6 digits

Heating method: PID temperature control constant temperature water bath

Heating power: 1000W, uniform heating

Nitrogen flow meter control: 0-3L/min

Temperature range: Room temperature -99.9 ℃

Determination lower limit: 16 mg/L

Temperature control accuracy: ± 1 ℃

Display mode: 7-inch LCD touch screen(with temperature display,Temperature control accuracy: ≤± 1%)

Nitrogen inlet pressure: 0.1Mpa

Nitrogen flow branch: 0.05-0.6L/min

Working power supply: AC (220 ± 22) V, 50HZ

Environmental temperature: (5-35) ℃(AC 220V,50HZ

Environmental humidity: (0-95)% RH