Multi functional distiller priceProduct Introduction:
The use of far-infrared heating method has the advantages of high thermal efficiency, long service life, fast temperature rise and cooling speed, adjustable heating time and power. The instrument can be equipped with an external water cooling circulation device, which introduces a constant flow of positive pressure gas during the distillation process, effectively improving the distillation speed and efficiency. The entire system is simple, easy to install and maintain, and reliable to use. It can be widely used in the pre-treatment process of sample analysis that requires distillation in industries such as environmental protection, chemical, pharmaceutical, and materials. Especially suitable for distillation treatment of samples before the determination of ammonia nitrogen, volatile phenols, cyanates, etc.
Multi functional distiller priceTechnical Indicators:
1. Can distill 5 samples simultaneously.
2. The heating power can be adjusted at intervals of 100 watts, with a maximum heating power of 1600 watts.
3. The heating time can be adjusted, with a maximum timer time of 2 hours and 30 minutes and a minimum timer time of 10 minutes.
4. Optional external water cooling circulation device, capable of cooling 5 condenser tubes simultaneously.
5. Built in gas generator, blowing air into 5 distillation bottles to improve distillation efficiency.
6. Power requirement: 220V (10A).
Usage (Taking Ammonia Nitrogen Distillation as an Example)
1. Accurately add 150mL of the sample into 5 distillation flasks and insert a reflux tube at the top of each flask. Ensure that the plastic king tube at the end of the reflux tube is at the bottom of the absorption bottle.
2. Accurately add 100mL of boric acid absorption solution into four receiving bottles, place the receiving bottles at the lower ends of the condenser tubes on both sides of the instrument, ensuring that the silicone rubber tube at the end of the condenser tube is below the liquid level of the absorption solution.
3. Turn on the cooling water and blowing power switch on the left side of the HCA-300 multifunctional distiller. Ensure smooth airflow in the distillation flask and water flow in the condenser tube.
4. Turn on the power switch of the electric furnace. The instrument displays a built-in heating power of "1000" watts. Press the "+" and "-" keys to adjust the heating power to "500" watts. After pressing the "timer" button, the instrument displays "00:00", and then press the "+" and "-" buttons to adjust the timer time to "00:20". At this point, it indicates that the scheduled time is 20 minutes. Afterwards, the instrument automatically heats up according to the set power timing until the heating is completed. Once the electric furnace reaches the heating time, the instrument automatically displays "OFF". At this point, although the heating is stopped, the cooling fan of the electric furnace continues to work to remove excess heat. If the heating power is changed during the heating process, the timing needs to be reset.
5. After removing the receiving bottle and diluting it to 250mL, it can be directly used for subsequent titration.
6. After the solution in the distillation flask has cooled to room temperature, turn off the condensate and blowing power switches on the left side of the instrument.
When conducting other sample analyses, the timing and heating power should be determined by the user based on experimental requirements and their own exploration.