Spectral ozone analyzer - Kipp Zonen
Brewer MkIII is a spectrophotometer used to measure ozone and sulfur dioxide in the atmosphere. It measures specific wavelengths of direct solar radiation to determine the total amount of ozone and sulfur dioxide columns in the atmosphere. It can also perform high-resolution ultraviolet spectral scanning of direct or total solar radiation. The software can accurately calculate UVA, UVB, UVE (erythema), and UVI (index).
Brewer MkIII spectrophotometer for ozone measurementEquipped with a spectrophotometer, installed on a dedicated azimuth tracker, tracking and measuring with solar or lunar light as the light source. Based on the absorption characteristics of ozone for UV, by accurately tracking the sun (or moon) and using diffraction spectroscopy technology, the light intensity of six wavelengths of UV (303.2 nm, 306.3 nm, 310.1 nm, 313.5 nm, 316.8 nm, 320.1 nm) is measured, and the total atmospheric ozone column is inverted based on the principle of differential absorption.
Brewer MkIII spectrophotometer for ozone measurementWidely applicable to a wide range of environments, it can be deployed for long-term outdoor monitoring. The spectrophotometer of the instrument can automatically compensate for the expansion and contraction of components caused by temperature changes. It can be used outdoors around the world without the need for complex temperature stabilization measures. At present, Brewer MkIII is used for monitoring work from the tropics to the north and south.
The MkIII 'Double' Brewer connects two photometers in series, which can improve the accuracy of ultraviolet light measurement in low sunlight conditions. The configuration of the MkIII 'Double' Brewer is significantly superior to the single spectrometer MkII Brewer and the discontinued MkIV Brewers.

Application field
Instrument approved by the World Meteorological Organization (WMO) for observing total ozone column
Provide measurement information for the World Ozone and Ultraviolet Data Center (WOUDC)
Automatic observation of total ozone column and sulfur dioxide
Study the Effects of Ultraviolet (UV) Radiation from the Sun and Sky
main advantages
The Brewer spectrometer is a recommended ozone column total observation instrument by the World Meteorological Organization.
The method of observing the total amount of ozone column using Brewer spectrometer is included in the meteorological industry standard of the People's Republic of China
Composed of a weather resistant spectrophotometer, azimuth tracker, and bracket
Near instantaneous observations of total ozone column, SO2, and ultraviolet spectra can be conducted
High wavelength accuracy
Suitable wavelength range
Reduce stray light
Ultra long stability
Automated celestial tracking
System Composition
Brewer spectrometer includes: spectrometer, control computer, and calibration system.
The spectrometer is the core component, mainly composed of a diffraction spectrometer, tracker, and tripod;
The calibration system mainly consists of standard lights and supporting facilities
Technical Specifications
Measuring principle |
Self compensating dual Ebert spectrometer |
UVMeasurement |
Direct solar radiation or total radiation, UVA, UVB, UVE, and UVI |
Sun tracker |
Integration, including heavy-duty tripod |
spectral range |
286.5 ~363 nm |
resolution |
0.6 nm±0.01 nm |
OzoneMeasure wavelength with SO ² |
303.2 nm• 306.3 nm • 310.1 nm 313.5 nm• 316.8 • nm • 320.1 nm |
Measurement uncertainty of chromatographic column ozone |
< 1 % |
Serial output |
RS-422 Comes with isolated RS-422 to RS-232C converter, AC power adapter, for connecting to PC's running software, PC included |
Operating temperature range |
-20 ~ 50℃ -50 ~ 50℃ (optional insulation cold cover) |
Storage temperature range |
-20 ~ 40℃ |
Power supply voltage |
115 Or 230 VAC, 50/60 Hz |
Detector type |
UVPhotomultiplier tube (PMT) |
GW-BASICsoftware |
BrewerOperation, data storage, and analysis |
Spectral ozone analyzer - Kipp Zonen