- Phone
-
Address
No.1 Shimei Ganyuan Industrial Road, Wanjiang, Dongguan City
Dongguan Jiasi Instrument Co., Ltd
No.1 Shimei Ganyuan Industrial Road, Wanjiang, Dongguan City
Shoe ozone aging test chamberMain technical parameters
Model: JS-CQ-150
| internal dimensions | W500*H600*D500mm |
| external dimensions | W950*H1800*D1000mm |
| temperature range | +10℃~80℃ |
| humidity range | 60%±5% |
| temperature deviation | ±2.0℃ |
| temperature fluctuation | ±0.5℃ |
| ozone concentration | 50~1000pphm (low concentration) |
| 50~190pphm (without humidity control under conditions) | |
| Ozone concentration deviation | ±10% |
| Sample rack lifting frequency | 1 R/min (adjustable) |
| airflow velocity | 12~16mm/s |
| Test gas flow rate | 20~70L/min |
| Sample rack speed | 360 degree rotating sample holder (adjustable speed of 1 revolution per minute) |
| test device | 3 sets of static and 3 sets of dynamic testing |
| Ozone concentration analysis | Concentration analysis regulator 4~20mA |
| thermal insulation material | PU rigid polyurethane foam, thickness 100mm |
| box door | Left side single door opening method, window frame/Two silicone rubber sealing strips and anti condensation electric heating devices are installed on the door frame, with observation windows of approximately W220xH280mm and lighting fixtures on top of the doors and windows |
| heating system | (1) Heater: High temperature insulated ceramic heating; (2) Control method: PID regulation, actuator: solid-state relay. |
| Shoe ozone aging test chamberFreezing system | (1) Working mode: mechanical refrigeration/air-cooled; (2) Control mode: The controller of the control system automatically adjusts according to the test conditions temperanceOperating conditions of the refrigeration unit; (3) Compressor: French Taikang high-efficiency fully enclosed compressor unit; (4) Evaporator: finned multi-stage automatic load capacity adjustment; (5) Condenser: Wind type condenser; (6) Expansion system: automatic capacity control refrigeration system; (7) Refrigerant: Environmentally friendly refrigerant R404 (with an ozone depletion index of 0)(8) Refrigeration process: nitrogen protection welding, two-stage rotary vane pump vacuum pumping to ensure Internal cleaning of refrigeration system, design of water tray at the bottom of compressor, condensation water Discharge to the outside of the box through the drainage pipe at the back of the box; (9) Refrigeration accessories: desiccant, oil separator, refrigerant flow window, repair Var, high voltage protection switch. |
| Ozone detection system
| (1) Ozone generator: using a voltage silent discharge tube to generate ozone; (2))Shoe ozone aging test chamberExhaust system: After the experiment is completed, the experiment will be conducted Ozone inside the box passes through the specializedUse exhaust fans and ducts to exhaust to the outside; |
| control system
| Programmable temperature and humidity controllerLCD touch controller, large LED display with high contrast and backlight function The interface can be operated in both Chinese and English to display the set test parameters, running time, heater, and refrigeration compressor. It also has the functions of automatic test operation and PID parameter self-tuning. The setting of parameters for system B adopts a human-machine dialogue method, only requiring the setting of temperature and humidity;The automatic operation function of the test box can be achieved. The C system uses intelligent control software to automatically combine cooling, heating, and other system conditions,Thus ensuring high-precision control over the entire temperature range, achieving the goal of energy conservation and consumption reduction. The detection device of D is capable of automatically displaying detailed faults and issuing alarms E-screen display: set temperature; Actual measured temperature; F Precision: Temperature: 0 1 ℃ humidity: 0.1% RH G-type capacity: 120 sets of 1200 segment programming, 999 cycle settings, maximum number of steps per program of 100, maximum time per program step: 99 hours and 59 minutes; Programs can be linked to each other; H line mode: Constant running, program running; |
| Safety protection device | 1. Heater surface temperature over temperature protection 2. Heater short circuit and overload protection device 3. Overheating protection for circulating motor c) High temperature and over temperature protection d) Reverse phase, under phase protection device, e) No fuse switch overload protection g) Line fuses and fully sheathed terminals |