Salt spray test chamber maintenance: We provide on-site repair services for salt spray test chambers. $r $n Maintenance/replacement of core parts: $r $n - spray system: unblock the clogged nozzle, calibrate the spray pressure and sedimentation (in accordance with GB/T2423.17 and other standards). $r $n-Temperature Control System: Repair heating tubes, temperature sensors, and temperature controllers to ensure stable temperature (such as 35 ℃ 2 ℃/47 ℃ 1 ℃). $r $n - Electrical system: Replace aging lines.

Product Details

Salt spray test chamber maintenanceIntroduction to on-site maintenance of Ningbo salt spray test chamber:

盐雾试验箱维修

The following are common faults and possible causes of salt spray test chambers, which can help you quickly troubleshoot problems:

🌫️ Abnormal spray volume

*Too small amount of spray or no spray

1. Improper spray regulator: the spray regulator is improperly placed or the pressure is set too low.

2. nozzle blockage: The salt spray nozzle is blocked by crystals or impurities.

3. Filter blockage: The glass filter at the air inlet is blocked, resulting in poor air flow.

4. Air source issue: The air compressor is not running or the main switch is not turned on.

🌡️ The temperature cannot reach the set value

*Slow temperature rise or inability to raise temperature

1. Setting error: The setting value of the temperature controller or safety protection switch is too low.

2. Heating system malfunction: Damaged heating tube, poor contact, or relay failure.

3. Poor sealing of the box: The sealing strip of the box door is aging and deformed, leading to heat leakage.

💧 Insufficient water level alarm

*Equipment alarm prompts water shortage

1. Low water level: The water level in the laboratory or saturated tank is below the normal standard.

2. Waterway issues: Blockage or leakage in the inlet pipeline.

🚫 Unable to spray

*The device is running but there is no salt spray generated

1. Solenoid valve fault: the solenoid valve controlling spray is damaged.

2. Nozzle blockage: The nozzle is blocked.

3. Pressure issue: Pressure gauge malfunction or insufficient air pressure.

⚙️ Unable to operate after the power is turned on

*Equipment unresponsive

1. Power failure: broken power cord, blown fuse, or tripped circuit breaker.

2. Water level protection: Some models of salt spray test chambers have water level protection function. If the water level in the heating water tank is too low, the power will be cut off to prevent dry burning.

Safety Reminder: During any inspection orSalt spray test chamber maintenanceBefore proceeding, please make sure to cut off the power supply of the equipment and wait for the temperature inside the box to drop to room temperature to prevent electric shock or burns. For complex faults involving circuit boards and core components, it is recommended to contact professionals for handling.