- Phone
-
Address
No. 748, Group 7, Wudun Village, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
Shanghai Juwei Instrument Equipment Co., Ltd
No. 748, Group 7, Wudun Village, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
Rapid temperature change test chamber 1000 litersJW-2009E-615-15
Rapid temperature change test chamber 1000 litersJW-2009E-615-15
Rapid temperature change test chamber 1000 litersJW-2009E-615-15
1. Sample Description |
||
|
This experimental equipment prohibits: Testing and storage of samples of flammable, explosive, and volatile substances Testing and storage of corrosive substance samples Testing or storage of biological samples Testing and Storage of Strong Electromagnetic Emission Source Samples |
||
2. Volume and size |
||
|
Nominal content volume Inner box dimensions Overall dimensionsabout |
1000L W1000×H1000×D1000 mm W1920×H2000×D 2450 mm(Height does not include motor and motor cover)
|
|
3. performance | ||
|
3.1.Test environment conditions 3.2. 3.3. 3.4.
3.5.test method |
The airflow around the equipment is unobstructed, without high concentration dust, corrosive or flammable gases The ambient temperature is5~35℃ relative humidity≤85%RH Circulating cooling water temperature≤30℃ The supply pressure of circulating cooling water is0.25~0.4MPa(water coolingThe device does not have this requirement The temperature change rate meets the relevant standards |
|
5.2.1 Temperature Range |
-60℃~+150℃ |
|
5.2.2 Temperature fluctuation |
≤ ±0.5℃ |
|
5.2.3 Temperature uniformity |
≤ 2℃ |
|
5.2.4 Temperature deviation |
≤ ±2.0℃ |
|
5.2.2 Temperature fluctuation |
≤ ±0.5℃ |
5.2.3 Temperature uniformity |
≤ 2℃ |
5.2.4 Temperature deviation |
≤ ±2.0℃ |
5.2.5 Temperature change rate:
heating rate:RT ~ +150℃ heating ratelinear15℃/min
降温速率:+70℃ ~ -55℃ 降温速率linear12℃/min
5.2.6 Humidity Range:Humidity range:20%~98%
5.2.7 Relative humidity deviation:+2~-3%(>75%RH)±5%(≤75%RH)
5.2.8 Humidity fluctuation:±2%RH