Jr-hda-vRapid hydrogen analyzerIt can accurately and quickly measure hydrogen in aluminum and aluminum alloy melts.

Aluminum hydrogen analyzerWorking principle:
Adopting the "closed loop circulation method", in accordance with Sievert's law (Alscan and Telegas hydrogen meters are strictly adhered to)Formula definition: [H]=K A K T S 0 pH 2
KA - correction factor for aluminum alloy *, related to alloy composition, pure aluminum is 1.00
PH2- hydrogen partial pressure in a closed loop (atmospheric pressure)
KTS0- Saturation solubility of hydrogen in aluminum under different temperature conditions, with a saturation solubility of 0.92ml/100g for pure aluminum at 700 ℃
Note: The JR-HDA-V hydrogen analyzer strictly conforms to the above formula, and its KA, KT, and S0 parameters all meet the general definition.
JR-HDA-V and ALSCAN conducted comparative tests on the same flow channel samples using their respective probes, and the errors were within a reasonable range (5% or ± 0.01ml/100g). Our company has on-site video materials available for those in need.
The differences in main functional configurations between JR-HDA-V and ALSCAN:
Motherboard: JR-HDA-V has better integration and anti-interference ability than ALSCNA
Sensor accuracy: The thermal sensor and the thermal sensor have comparable accuracy over the full range
Upper computer software: capable of graphically reading measurement files from instruments, used for data analysis and management, as well as remote computer control of instruments
Display: JR-HDA-V graphical interface, superior to ALSCNA digital tube, convenient for on-site control of sampling probe operating conditions.
USB data management: JR-HDA-V can transfer measurement files to personal computers for data analysis and management anytime and anywhere
Automatic termination after measuring balance: Extend probe life while ensuring accurate data conditions
Power supply: JR-HDA-V is equipped with a 9.8Ah lithium battery and built-in charger, with a theoretical working time of over 7 hours
Humidity hydrogen content: equipped with a humidity meter to measure the environment (the instrument automatically calculates the hydrogen content of the aluminum melt in equilibrium with the environmental humidity)
Probe failure detection: warning prompt when the measurement curve does not conform to the defined measurement principle
Measurement balance detection: There is a warning prompt when the measurement curve does not conform to the defined measurement principle
Data storage: JR-HDA-V stores 500 sets of measurement data in file format, while ALSCNA stores over 100 sets
Probe: The service life of both hydrogen measuring probes is equivalent, and the measurement reproducibility of the same sample with different probes is good (within an acceptable range of error)
On site printing: Real time printing of measurement process and result data
Long cycle measurement: facilitates the adjustment of various process parameters of the degassing equipment during process research (JR-HDA-V long cycle measurement is 60 minutes)
Communication: Used for connecting with personal computers, remote control operations, and transferring measurement files to personal computers for data analysis and management after connection
Probe swing: used when necessary to accelerate the measurement speed of molten materials that are stationary or have a flow rate less than 5cm/s
Gas cylinder capacity: JR-HDA-V supports more data in one inflation, approximately 150 measurements, with the same inflation method for both
There is video evidence to support this: HDA-V has conducted on-site comparative testing videos with ALSCNA hydrogen meters in six companies. Leakage experiment video of external circulation probe (aluminum oxide hydrogen probe)
HDA-V hydrogen analyzer is equipped with HP-xxxxx series hydrogen probe, which can also be used for ALSCNAHydrogen analyzer, HP-xxxxx series hydrogen measuring probe is an internal circulation probe, and its service life is similar to ALSCAN hydrogen measuring probeThe probe is equivalent (see technical specifications in the table below) to the external circulation probe (used with the ELH hydrogen analyzer for oxidation)Compared with aluminum hydrogen probes, aluminum hydrogen probes have the following characteristics:
1. The internal circulation probe ensures no "leakage" and its measurement results are accurate and true. Existing external circulationThe probe generally has a problem of "leakage" (which cannot be overcome by existing technology), and its measurement results are low.
2. Due to their different working principles, the "attenuation failure" effect of the internal circulation probe is significantly better than that of the external circulation probeThe ring probe has a longer service life than existing external circulation probes.
3. The internal circulation probe is not affected by the melt flow rate or probe insertion angle, therefore the measurement accuracy is higher,More convenient operation.
4. The connection between the internal circulation probe sampling head (part in the aluminum melt) and the stainless steel capillary tube is integratedSeamless connection, capable of withstanding high temperatures (800 ℃), while existing external circulation probes are made of epoxy resinSealing (* can withstand up to 200 ℃). Therefore, in harsh on-site environments, the external circulation probe is capable ofEasy to leak and fail.
5. The external circulation probe is prone to splashing of molten material during the measurement process (in a blown state), causing safety hazards to both personnel and the instrument.
Jr-hda-vRapid hydrogen analyzerMain technical indicators:
Measurement range: 09.99ml/100g
Calibration range of standard sample: (0-92)% hydrogen standard gas
Resolution: 0.001 ml/100g
Measurement accuracy: ± 0.01 ml/100g or ± 5% [H], whichever is higher
Measurement time (automatic): Typically within 5-15 minutes, manual measurement can be set for 60 minutes
Display: 640 × 480 (5-inch touch color screen) LCD dot matrix display
Printer: Thermal dot matrix printing
Data storage: 512Mbitye data storage, capable of storing 100 alloy numbers and their correction coefficients, 5000 sets of hydrogen content data (measured every 20 seconds)
Melt temperature measurement: K-scale thermocouple, capable of continuous long-term measurement in aluminum liquid, with an accuracy of ± 3 ℃
Communication: Standard RS-485 interface, capable of remote control of instruments through upper computer software
Data transfer: standard USB interface
Nitrogen filling cylinder: 700ml capacity, high working pressure of 10Mpa, capable of approximately 150 measurements
Average lifespan of the probe: 10 insertions into aluminum liquid or 3 hours of cumulative circulation in pure aluminum liquid, whichever comes first
Environmental temperature: 1060 ℃
Power condition: 170264 VAC 47 63Hz or 12VDC lithium battery pack, 9.8Ah. Can work continuously for about 5 hours
Dimensions and weight:
仪器: 360mmL×210mmW×280mmH14.5kg
Calibration attachment: 200mmH × 120mmW × 140mmD1kg
Probe length: 1100mmL
Supply scope of JR-HDA-V hydrogen analyzer:
1. One host
2. One instrument operation manual
3. 10 probes
4. Two electric couples
5. One electronic version of the instrument user manual
6. Measurement arm set
The storage format for JR-HDA-IV measurement files is as follows:
Instrument number: 01234567
Software version number: Ver1.0.1
Calibration date and time: 2009/05/11 17:25
Standard curve: V0:1003 V1:5036 V2:12520 X1:20.45% X2:81.00
Measurement number: T0000185
Sampling method and measurement group: Automatic (2/3) balancing 3 times
Alloy number: 1235
Alloy correction factor: 1.00
Alarm upper limit: 0.30 (mL/100g)
Alarm lower limit: 0.07 (mL/100g)
Environmental temperature:----
Environmental humidity:-----
Sensor zero position value: 01103-----
Start measurement date and time: 2009/05/20 15:07:20
Measurement end date and time: 2009/05/20 15:16:20
Warning or prompt---------------
----Measurement completed----
[H]:0.292 (mL/100g)
PH2:0.05460 atm
C[H]:0.000 (mL/100g)
T:748 (℃)
Measurement data recording every 20 seconds:
[H] PH2 C[H] T R t
(mL/100g) (atm) (mL/100g) (℃) (bt) (s)
0.116 0.00864 0.000 747 01292 0020
0.151 0.01481 0.000 747 01427 0040
0.179 0.02062 0.000 747 01535 0060
0.198 0.02538 0.000 747 01623 0080
0.213 0.02933 0.000 747 01696 0100
0.225 0.03269 0.000 747 01758 0120
0.236 0.03561 0.000 748 01812 0140
0.244 0.03810 0.000 748 01858 0160
0.251 0.04037 0.000 748 01900 0180
0.258 0.04248 0.000 748 01939 0200
0.265 0.04432 0.000 749 01973 0220
0.270 0.04595 0.000 749 02003 0240
0.272 0.04730 0.000 748 02028 0260
0.277 0.04844 0.000 749 02049 0280
0.278 0.04941 0.000 748 02067 0300
0.280 0.05017 0.000 748 02081 0320
0.282 0.05076 0.000 748 02092 0340
0.283 0.05136 0.000 748 02103 0360
0.285 0.05201 0.000 748 02115 0380
0.289 0.05266 0.000 749 02127 0400
0.290 0.05320 0.000 749 02137 0420
0.290 0.05368 0.000 748 02146 0440
0.291 0.05412 0.000 748 02154 0460
0.292 0.05444 0.000 748 02160 0480
0.292 0.05460 0.000 748 02163 0500
0.293 0.05471 0.000 748 02165 0520
0.292 0.05460 0.000 748 02163 0540

Instrument number: 01234567
Software version number: Ver1.0.1
Calibration date and time: 2009/03/10 14:47
Standard curve: V0:1003 V1:5036 V2:12520 X1:20.45% X2:81.00
Measurement number: T0000024
Sampling method and measurement group: Automatic (2/3) balancing 3 times
Alloy number: 1235
Alloy correction factor: 1.00
Alarm upper limit: 0.32 (mL/100g)
Alarm lower limit: 0.08 (mL/100g)
Environmental temperature: 28 ℃
Environmental humidity: 40% RH
Sensor zero position value: 01782-----
Start measurement date and time: 2009/03/27 11:54:15
Measurement end date and time: 2009/03/27 12:06:35
Warning or prompt---------------
----Measurement completed----
[H]:0.173 (mL/100g)
PH2:0.03395 atm
C[H]:0.115 (mL/100g)
T:703 (℃)
Measurement data recording every 20 seconds:
[H] PH2 C[H] T R t
(mL/100g) (atm) (mL/100g) (℃) (bt) (s)
0.071 0.00574 0.115 703 01908 0020
0.092 0.00966 0.115 702 01994 0040
0.104 0.01235 0.115 703 02053 0060
0.112 0.01431 0.114 703 02096 0080
0.119 0.01606 0.115 703 02131 0100
0.124 0.01756 0.115 703 02159 0120
0.129 0.01884 0.115 703 02183 0140
0.132 0.01991 0.115 703 02203 0160
0.136 0.02087 0.115 703 02221 0180
0.139 0.02177 0.115 703 02238 0200
0.141 0.02263 0.115 703 02254 0220
0.143 0.02338 0.114 702 02268 0240
0.145 0.02412 0.114 702 02282 0260
0.147 0.02482 0.114 702 02295 0280
0.149 0.02541 0.114 702 02306 0300
0.150 0.02605 0.114 702 02318 0320
0.152 0.02663 0.114 702 02329 0340
0.155 0.02722 0.115 703 02340 0360
0.156 0.02775 0.115 703 02350 0380
0.157 0.02834 0.114 702 02361 0400
0.160 0.02888 0.114 703 02371 0420
0.160 0.02936 0.115 702 02380 0440
0.162 0.02984 0.115 703 02389 0460
0.163 0.03032 0.115 703 02398 0480
0.165 0.03074 0.115 703 02406 0500
0.164 0.03112 0.115 702 02413 0520
0.165 0.03144 0.114 702 02419 0540
0.167 0.03170 0.115 703 02424 0560
0.168 0.03197 0.115 703 02429 0580
0.168 0.03218 0.115 703 02433 0600
0.169 0.03245 0.115 703 02438 0620
0.170 0.03272 0.115 703 02443 0640
0.169 0.03299 0.115 702 02448 0660
0.171 0.03320 0.114 703 02452 0680
0.172 0.03347 0.115 703 02457 0700
0.172 0.03368 0.115 703 02461 0720
0.173 0.03395 0.115 703 02466 0740

Technical Services:
Our company provides free technical training services for the operators of the demand side, and promises to waive the fees for the instruments sold within one yearFee warranty. During the warranty period, if the instrument is damaged due to non-human or improper use, our company will provide free repairs.
Standard sample purchasing agent:
Regarding the procurement of standard samples, our company suggests purchasing them locally. If the customer does not have them locally, our companyIt can be purchased on behalf of others in Chengdu. Generally, it is recommended to purchase the following standard gases:
Hydrogen content in N2 (volume fraction): 2% per bottle (8-liter aluminum bottle, 10Mpa) and pressure reducing sampling valve
Hydrogen content in N2 (volume fraction): (20%) One bottle (8 liters aluminum bottle, 10Mpa) and reduced pressure