Rack mounted hydrogen zero one unit machineTechnical specifications
Zero gas flow rate: 0~5000 mL/min
Zero air pressure: 0.1-0.6 MPa (without built-in air compressor, external air pump as feed gas)
System pressure: 0.6-0.8 MPa
Voltage stabilization accuracy: 0.003 MPa
Zero dew point:<-20 ℃
Total hydrocarbon and methane content:<0.05 ppm
Output particles:<0.01 μ m
Hydrogen purity: 99.999%
At a dew point of -56 ℃, the oxygen content is less than 3ppm
Hydrogen flow rate: 0-300/500 mL/min
Traffic display: LED digital display
Work pressure: 0.4 MPa (adjustable by contacting the manufacturer)
Voltage stabilization accuracy: 0.001 MPa
Rack mounted hydrogen zero one unit machinecharacteristic
1. Built in single-chip computer, intelligent management of hydrogen generation and zero gas treatment, fully automatic operation, automatic intelligent fluid replenishment, safe and reliable.
2. Select imported special precious metal electrolytic materials to effectively improve electrolysis efficiency, maintain a constant temperature of the electrolytic cell body, and extend the service life of the electrolytic cell.
3. The gas path is entirely made of stainless steel pipes (electrolytic polishing, ultrasonic cleaning), with multi-stage water and dust removal to effectively ensure the quality of the exhaust gas.
4. The dehydration section is activated alternately with Nafion and silica gel to remove water to the maximum extent possible, and there is a trace amount of oxygen removal agent at the outlet (no activation required)
Make the purity higher.
5. * The anti backflow device ensures the phenomenon of instrument backflow.
6. The instrument design is compact and flexible, with an external air pump unit to save space.
7. The zero gas section adopts an imported catalyst skeleton and a dehydration and dust removal method, which prolongs the lifespan and improves the efficiency of the catalyst.
8. Built in intelligent protection measures, capable of protecting against overheating, overpressure, and overcurrent, high-precision voltage stabilization, stainless steel gas storage tanks, and more