RTC-156 can provide a wide temperature range: from -30 ℃ to 155 ℃. A dry furnace can cover a temperature field span of 185 ℃.
1.RTC dry furnace has obvious advantages in temperature rise and fall speed. The heating and cooling speed has increased by 20% compared to JOFRA's old products, resulting in a significant reduction in equipment downtime and calibration costs
2.DLC dynamic load compensation system (patent application pending) ensures excellent temperature field consistency of the casing
3.The reference probe is equipped with an intelligent chip that can achieve plug and play functionality
4.Can be done throughUSB interface and computer communication
5.colorVGA LCD display screen for easy reading and monitoring of temperature field status
6.Intuitive, fast, and user-friendly interface
7.Lightweight and easy to carry
8.New multifunctional portable aluminum box
9.The new porous sleeve kit is compatible with almost all sizes of sensors
1.Temperature range:-30~155℃;
2.Accuracy:±0.1℃(Built in standards,type a),±0.04℃(join in marriageExternal standard,OnlyBC type);
3.Stability:±0.005℃;
4.Display resolution:1/0.1/0.01/0.001 ℃ adjustable;
5.Calibration chamber (calibration sleeve) specifications(mm):Φ30 × L150;
6.Radial consistency:0.01℃;
7.Heating time (from ambient temperature to maximum temperature):15minute;
8.Cooling time (from maximum temperature to)ambient temperature):12minute;
9. Automatic stepping: maximum20 steps, adjustable holding time for each point, calibration results can be stored;
10.Maximum power:400W;
11.Instrument size (length)X Width x Height):362×171×363mm;
12. Weight:10.5kg。
The following parameters are limited totype B
13.Current measuring range/Accuracy: 0~24mA/±(0.005%Rdg+0.01%F.S);
14. Voltage measurement range/Accuracy: 0~12V/±(0.005%Rdg+0.01%F.S);
15.RTD resistance measurement range/accuracy: 0~4000Ω/±(0.005%Rdg+0.005%F.S);
16.RTD measurement type: Pt10/50/100/200/500/1000, Cu50/100, ± 0026℃@0℃(Pt100);
17.thermocoupleMV measurement range/accuracy: ± 78mV/± (0.0)05%Rdg+0.005%F.S);
18.Thermocouple temperature measurement type/Accuracy: E/J/K/T/R/S/B/N ± 0.11 ℃@320℃(K);
19.Cold end compensation range/Accuracy: 0~40 ℃/± 0.3 ℃
20.Switch test: dry contact5V,2.5mA;
21.24V circuit power supply: 24VDC ± 10%, maximum 28mA.
The following functional parameters are limitedB. Type C
22.External reference input type/Accuracy: Four wire Pt100400 Ω/± (0.0012% Rdg.+0.0005% F.S.) ± 0.008 ℃ @ 0 ℃;
DLC dynamic load compensation: thermocouple multi-point probe, temperature difference measurement.