RG-600P Fourier Portable Gas AnalyzerIt is a portable Fourier transform infrared gas analyzer independently developed by our company for environmental monitoring and industrial site emission gas analysis. This analyzer can measure the concentration of most gas pollutants and has the characteristics of high measurement accuracy, high stability and reliability, and fast response time.
RG-600P Fourier Portable Gas AnalyzerApplication scenarios:
Garbage incineration monitoring (analyzing SO)2、NOx、CO、HCl、HF、NH3Etc.)
Greenhouse gas monitoring (analyzing CO)2、N2O、CH4Waiting for gas concentration
Concentration detection of hydrocarbon components
Motor vehicle exhaust emission testing (analysis of NO, NO)2、CO、CO2、HC)
Measurement of standard gas substance concentration, detection of impurity concentration in standard gas substance
Product Features
Independently developed and designed interferometers for on-site applications, sturdy, durable, and maintenance free for life;
The interference part uses a helium neon laser, which has higher wavelength stability compared to semiconductor lasers;
All metal multiple reflection high-temperature gold-plated gas chamber, corrosion-resistant, highly reliable, and water-soluble gas components such as HCl and NH3Waiting for the loss of no components;
The resolution of the spectrometer is optimized for the gas under test, resulting in high detection accuracy;
The instrument has high internal temperature control accuracy and good environmental adaptability;
Adopting a long-life infrared light source;
Modular design is adopted for light sources, spectrometers, core circuits, gas chambers, etc., with high reliability, good scalability, and easy maintenance.
Sampling probe, sampling pipeline
High temperature probes are suitable for sampling gases containing dust or suspended particles, especially for measurement under non static conditions;
Contains a non corrosive, heated, replaceable filter element;
This filter element is installed in an electrically heated stainless steel filter housing with thermal insulation There is a protective outer box;
Temperature regulation is controlled by PTC heaters;
Connect the heated sample wire directly to the probe housing with a quick connector
High temperature pretreatment
Small size and light weight, convenient to carry;
Automatic temperature control, with a control accuracy of 0.1 ℃;
Heating fault automatically stops sampling to prevent system contamination;
Ensure high-precision filter (0.1u) and system pressure drop of 6mbar. ·Humanized design, easy replacement of filters;

project |
Specifications |
Instructions |
SO2measurement range |
(0~50~2000)mg/m3 |
Scalable range |
NO measurement range |
(0~150~2000)mg/m3 |
Scalable range |
NO2measurement range |
(0~150~2000)mg/m3 |
Scalable range |
N2O Measurement range |
(0~50~2000)mg/m3 |
|
CO measurement range |
(0~75~2000)mg/m3 |
Scalable range |
CO2measurement range |
(0~25~50)% |
Scalable range |
HCl measurement range |
(0~75~200)mg/m3 |
Scalable range |
HF measurement range |
(0~50~200)mg/m3 |
Scalable range |
NH3measurement range |
(0~50~200)mg/m3 |
Scalable range |
CH₄measurement range |
(0~50~2000)mg/m3 |
Scalable range |
H2O Measurement range |
(0~40)% |
|
detection limit |
≤1% F.S. |
|
indication error |
≤2% F.S. |
|
repetitiveness |
Not exceeding 1% |
|
stability |
Not more than 2% |
|
Suggested traffic |
(2L~3L)/min |
|
24-hour zero drift |
Not exceeding ± 2% FS |
|
24-hour range drift |
Not exceeding ± 2% FS |
|
response time |
not greater than 100s@3L /min |
HCl and HF should not exceed 200 seconds |
Gas chamber temperature |
180℃ |
|
Working temperature range |
(0~40)℃ |
|
ambient humidity |
≤98% |
|
Power Supply/Power |
220VAC ± 10%/peak 2500W Average 150W @ ambient temperature 25 ℃ |
|
Specification of pneumatic joint |
Φ 6mm card sleeve |
air intake |
Φ 8mm card sleeve |
air vent |
|
Φ 6mm quick twist |
Optical path blowing |
|
warm-up time |