E-mail
ruixinjie01@163.com
-
Phone
18611697209
-
Address
B1-1514, Xinlong Building, No. 87 Building Materials City West Road, Changping District, Beijing
Beijing Ruixinjie Environmental Protection Technology Co., Ltd
ruixinjie01@163.com
18611697209
B1-1514, Xinlong Building, No. 87 Building Materials City West Road, Changping District, Beijing
MCM40K3 manual quality comparator:
Order number MCM40K3 with uncalibrated weather sensor
Order number MCM40K3 DAkS with calibrated weather sensor and DAkS certificate
Model 6
Maximum range 41 kg
Readability 1 mg
Application range 0-41 kg
Repetitive s
Under optimal conditions, 1): 2mg
Under standard conditions, 2): 3 mg
1/3 load 2)
1/10 load 2): 2 mg
Under standard conditions, F3: 6 mg
inElectronic weighing | Peeling range: 41 kg
Linear: 20 mg
Off load error: 3.5 mg/mm
Stable time: 5s
Cycle time (ABA): 120s
MCM40K3 manual quality comparator standard attachment
Data interface: RS-232C, USB, Ethernet, SD card (optional RS-232C, PS2, Bluetooth)
Other applications: weighing, unit conversion, identifier, density measurement, statistics
Hook weighing equipment: with optional accessory 69EA0040
Comes with optional accessory 69EA0040
Comes with optional accessory 69EA0040
Meteorological sensor: can be connected to an external PC
Optional attachments
Calibration weight: 20 kg | E2
YCW722-00
Meteorological module: YMC20MC
Calibrated meteorological module: YMC20MC DAkkS
Second Windproof Cover: YDS05C | YDS03C
Upgrade equipment for 10kg: YAW51
Upgrade equipment for 20kg: YAW52
Lift equipment for 50kg
Lift with a chain hoist
Grab weight pliers with handle weights
Stainless steel column installed on the ground
size
Weighing tray size (width x depth): 400x300mm
Weighing room (width x depth x height): 400x 326x126 mm
Electronic device (width x depth x height): 239x 320x56 mm
Repeatability, also known as standard deviation's'; It is calculated after 5 ABA cycles and eliminating drift
1) Optimal conditions: In the E1 quality standard laboratory, a stone made shock-absorbing workbench is used for automatic measurement without being affected by the operator, under the condition of no wind on the upper part
2) Standard condition E: In the E1 quality standard laboratory, manually measure on a stone seismic workbench with no wind on the upper part
3) Standard condition F: In a laboratory that meets at least F1 standards, manually measure on a stone shock resistant workbench with air conditioning and a gentle breeze on the upper part